The Clinispin horizon 853VES centrifuge is capable of storing up to 10 user-defined presets.  

These memory presets contain all of the settings needed to define a specific run, (run time, speed, 

acceleration rate, braking rate, etc.). The user can use these memory locations to quickly configure 

the centrifuge for a specific test type and ensure that the test is run in the same way each time by 

recalling this same setting.

Store a configuration in a memory location

 

1. 

Adjust the various settings (speed, time, acceleration rate, etc.) for the    

 

 

 

configuration to be saved.

 

2. 

Press and hold the PROGRAM button until the preset display begins to flash.

 

3. 

Press the MEMORY button to increment the memory location by one. Continue    

 

 

to press the memory button until the desired location is displayed.

 

4. 

Press the program button to store the current configuration into the memory  

 

 

 

location that is displayed.

Recall a configuration from a memory location

 

1. 

Press the MEMORY button until the desired memory location is displayed in the    

 

 

memory display.

 

2. 

As you review the presets, the settings on the control panel will change to  

 

 

 

reflect the settings stored in the memory preset currently displayed.

 

3. 

When the desired preset is found, pause and the memory display will    

 

 

 

momentarily flash. The stored configuration is now loaded.

notes:

Once a configuration has been loaded, it will remain active until any change is made on the control 

panel.  If for instance the user adjusts the time, the current configuration will no longer match the 

previously loaded configuration and the memory display will change to a “-” dash. The previous 

configuration may be recalled using the procedure described above.

Memory locations my be overwritten by selecting them during the store procedure.

digiTal cyclE counTEr:

This centrifuge is equipped with a digital cycle counter to monitor machine usage and for routine 

maintenance.

To display the current cycle count, perform the following:

 

1. 

Unlock the lid by pressing the OPEN/STOP button.

 

2. 

Open the lid latch so that the yellow LOCKED indicator light is not illuminated.

 

3. 

Press and hold the START button for approximately five seconds. The word  

 

 

 

“CYCLES” will be displayed in the speed display and the current cycle count will    

 

 

be displayed in the time display. If the current cycle count exceeds 9,999  

 

 

 

cycles, the cycle count will be displayed by showing the first part of the number    

 

 

and then the end of the number. For example: the 1st shown is 0002 and the     

 

 

2nd shown is 4537 represents 24,537 cycles.

MEMory locaTions / sToring and rEcalling:

page 12

Содержание 7711039

Страница 1: ...clinispin horizon 853VES Laboratory Centrifuge Rev 1 4 Operator s Manual P N 7711039 ...

Страница 2: ...unctions Time Adjustment and Timer Operation page 10 Speed Adjustment RCF Display page 10 Starting and Stopping a Run page 10 Unlocking the Lid page 10 Advanced User Settings Acceleration Rate page 10 Braking Rate page 11 Countdown Delay page 11 Imbalance Detection page 11 Audible Setting page 11 Memory Locations Storing and Recalling page 12 Digital Cycle Counter page 12 Care and Preventative Mai...

Страница 3: ...d force capability produces platelet poor free plasma and greatly reduces processing times The exclusive Woodley horizontal rotor allows for quick and easy sample loading and complete horizontal separation Save settings in one of 10 memory locations and recall them at the touch of a button Supplied Accessories Standard See next page for optional rotor accessories Also included not shown One 1 10 M...

Страница 4: ...32 100 mm tube holder p n 7713040 75 mm tube holder p n 7713043 250mm Cushion Spacer p n 1525 Part No Max Tube Size Speed Range Force Range 12 Place Performance Plus Horizontal 7786024 Comes with Twelve 12 of each of the tube holders listed 75mm tube holder 7713043 12 17mm x 75mm 500 to 5 000 RPM 40 to 3 100 xg 100mm tube holder 7713040 12 17mm x 100mm 500 to 5 000 RPM 40 to 3 800 xg 12 Place Hori...

Страница 5: ...n RPM Change Buttons RPM Display Running Indicator Light Locked Indicator Light Unlocked Indicator Light Start Button Memory Button Memory Preset Display Program Button Up Down Time Change Buttons Time Display RCF Conversion Button Control Panel see below Cabinet Open Stop Button Brake Display ...

Страница 6: ...e properly grounded 2 For operator safety the locking system is always active requiring power and direction from the user to disengage it the lid also automatically unlocks at the end of a run when it is safe to do so To unlock the lid in order to access the rotor chamber press the OPEN STOP button on the control panel The UNLOCKED indicator light should illuminate If it does not refer to the sect...

Страница 7: ... 11 for a description of these features 7 Begin the run by pressing the START button on the control panel 8 The centrifuge should begin to spin The RUNNING indicator light should illuminate IF A PROBLEM IS FOUND DURING A SPIN THAT REQUIRES THE CENTRIFUGE TO SHUT DOWN PRESS THE OPEN STOP BUTTON IMMEDIATELY 9 After time has elapsed the RUNNING indicator light will extinguish and the rotor will decel...

Страница 8: ...to hold tubes up to 10mL up to 17mm x 100mm The blue holders are de signed to hold tubes up to 5mL up to 17mm x 75mm Your centrifuge must contain a balanced load in order to work properly To ensure that the load is balanced keep these rules in mind when inserting test tube samples into the 12 place horizontal rotor DIRECTIONS 1 Compare the tube to be spun with the three boxes shown below 2 Find th...

Страница 9: ...inserting test tube samples into the 20 place fixed angle rotor 1 Opposing tube holders must be identical 2 Opposing tube holders must be empty or loaded equally weighted samples 3 If an odd number of samples is to be spun use a water filled tube to mate with the unpaired one DIRECTIONS 1 Compare the tube to be spun with the four boxes shown below 2 Find the box that most closely matches the tube ...

Страница 10: ...f the square of the speed this jump size will vary Starting and Stopping a Run With the lid switch closed press the START button to begin a run Press the STOP OPEN button at any time during a run to terminate it Unlocking the Lid The lid is unlocked whenever the red UNLOCKED indicator light is illuminated The lid unlocks automatically for 60 seconds at the end of a run The lid can also be unlocked...

Страница 11: ...s equipped with imbalance detection This ensures that the centrifuge will safely shut down in the event that an imbalanced load is run or a malfunction occurs This feature is intended to provide for operator safety and to extend the life of the centrifuge The LOW setting will allow for a greater imbalance before shutting the centrifuge down 1 While the centrifuge is idle press the PROGRAM button 2...

Страница 12: ...yed 3 When the desired preset is found pause and the memory display will momentarily flash The stored configuration is now loaded Notes Once a configuration has been loaded it will remain active until any change is made on the control panel If for instance the user adjusts the time the current configuration will no longer match the previously loaded configuration and the memory display will change...

Страница 13: ...aining fragments may provide stress points on subsequent tubes and result in additional breakage If a tube breakage occurs carefully remove the tube holder Properly dispose of the sample and tube fragments and thoroughly clean both the inside and outside of the tube carrier holder Insert a new tube cushion if necessary and replace the tube carrier holder in the rotor 4 Motor and Electrical Mainten...

Страница 14: ...owing table for fault information ERROR SPEED The centrifuge cannot reach full speed due to a problem with the rotor an inadequate power supply or other electrical problems Press the OPEN STOP button to cancel the error and then check the rotor and the line voltage Contact your authorized dealer or Woodley Equipment Company for further assistance ERROR BALANC The centrifuge has detected an imbalan...

Страница 15: ... lid knob enough to cause the pawl to lose contact with the lid safety switch If this happens the centrifuge motor may lose power but the lid will still remain locked If the knob is accidentally moved and this situation should occur rotate the knob fully clockwise to its stop position and the centrifuge will resume operation Circuit Breaker The Horizon is protected with a 4 Amp circuit breaker loc...

Страница 16: ...t parts or accessories 125 mm Tube Holder p n 7713044 Tube Holder Cap p n 7713011 75 mm Tube Holder p n 7713043 100 mm Tube Holder p n 7713040 20 Place Fixed Angle Rotor p n 7786042 50mL tube holder fixed angle p n 7787051 6 Place 50mL Fixed Angle Rotor p n 7786044 250 mm Tube Cushion p n 1525 6mm Tube Cushion p n 9150 Tube Holder Cap p n 7713011 75 mm Tube Holder p n 7713043 100 mm Tube Holder p ...

Отзывы: