User’s Manual – PGT
®
130.DT
Part No.: 7100.PGT130.DT
Page 13 / 21
V12/2020
Special Function:
Service-access without measurement validation
Users whose names are marked with a leading asterisk "*" are allowed to pass even if the
measurement values are out of range. The signal will indicate “yellow” instead of “green” and no
label will be printed.
Firmware Update
Use
Configuration -> Firmware upload -> Search
to select the appropriate BIN file. Only newer
versions can be installed, downgrade is not possible. Use the
Save settings
button on the bottom of
the page and wait until the update has finished (at least 3 minutes).
Do not disconnect the power supply during the update.
Operation
Test with user identification
If user identification is enabled the measure profile is controlled by the field “profile” of the user
record. Device default settings are used if no user profile is present.
1. Identify yourself with a registered RFID card.
2. Arrow indicators show the required measurement according to the specified user profile.
Left – wrist strap / right - shoes.
3. Touch only one sensor plate for measurement.
Test without user identification
If user identification is disabled the device settings are used.
Touch the left sensor for wrist measurement, or the right sensor for shoe- or combined
measurement.
Modes / Profiles
Wrist test
Put on the Wrist strap / smock with coil cord and connect the coil cord to the appropriate socket on
the left side of the PGT
®
130.DT. Touch the left sensor. After a short measurement time the color test
result (green= pass, / red = fail) and the measurement values are displayed.
Shoe test
Stand on the footplate with both shoes.
Touch the right electrode. After a short measurement time the color test result (green= pass, / red =
fail) and the measurement values are displayed.
Shoe and wrist test
Put on the Wrist strap / smock with coil cord and connect the coil cord to the appropriate socket on
the left side of the PGT
®
130.DT. Stand on the footplate with both shoes.
Touch the electrode. After a short measurement time the color test result (green= pass, / red = fail)
and the measurement values are displayed.