![Witeg WITOPET 5 402 002 Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/witeg/witopet-5-402-002/witopet-5-402-002_manual_989869017.webp)
17
SPECIFICATIONS
Model
Artivle no.
Volume
R[%]
VK[%]
PW2
5 402 002
Min
Max
0,2
1,0
2,0
12,0
2,7
1,5
6,0
1,3
0,7
PW10
5 402 010
Min
Max
0,5
5,0
10,0
4,0
1,0
0,5
4,0
0,8
0,4
PW20
5 402 020
Min
Max
2
10
20
4,0
1,0
0,8
3,0
0,7
0,4
PW100
5 402 100
Min
Max
10
50
100
1,6
1,0
0,8
0,8
0,4
0,3
PW200
5 402 200
Min
Max
20
100
200
1,2
1,0
0,6
0,6
0,3
0,3
PW1000 5 402 901
Min
Max
100
500
1000
0,9
0,7
0,6
0,50
0,50
0,50
PW5000 5 402 950
Min
Max
1000
2500
5000
0,6
0,6
0,5
0,3
0,3
0,2
These specifications are obtained in forward mode, using a gravimetric method with the
temperature of the distilled water, tips and all other conditions stabilised between 19°C and 21°C.
Number of measurements - minimum 15. The values given include all components of error due to
both normal handwarming and the changing of the tip.
These specifications were obtained using tips of best quality.
Performance tests : The specifications for the volumetric performance of the pipette are derived
from extensive gravimetric tests as prescribed by the Deutsches Institut für Normung e.V. To
verify the performance of your pipette, request document DIN 12650.