19 

3)  Place the head-half of the base on the cross brace and slot it into the foot-half. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

4)  Tighten the rondo screws 

 

5)  Connect the variable height jack, leg rest jack (depending 

on option) and plug into the mains 
 
 

7.2. Dismantling 

 

1)  Brake the wheels and place the bed base in low position

.

 

 
 
 
 
 
 
 
 

2)  Unplug from the mains and disconnect the variable height jack and leg rest jack (depending on  

option). (See stage instructions for dismantling the motors) 

3)  Loosen the two Rondo screws

 connecting the two halves of the bed base by 1cm and remove the 

clip. 

 

4)  Remove the head-half of the 

bed base. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

5)  Remove the clips  

and the foot-half of the bed base. 
 
 
 

 

 

Содержание AERYS 120

Страница 1: ...9_04_01 580046 Anglais 4 Le Pas du Ch teau 85670 SAINT PAUL MONT PENIT TEL 33 0 2 51 98 55 64 FAX 33 0 2 51 98 59 07 Email info medicatlantic fr Site Internet http www winncare fr DUO DIVISYS DO8C P 8...

Страница 2: ...6 4 3 NOISE 14 6 5 ELECTRICAL CONNECTION DIAGRAM 15 6 5 1 DOUBLE ELECTRIC STATEMENTS FOLDER DOUBLE STATEMENTS ELECTRIC FOLDING LEGS WITH KNEES 15 6 5 2 DOUBLE ELECTRIC STATEMENTS FOLDER DOUBLE STATEM...

Страница 3: ...FTING POLE AND IV STAND 25 10 MAINTENANCE 26 10 1 IDENTIFICATION 26 10 2 INSTRUCTIONS FOR DISMANTLING THE MOTORS 26 10 3 MAINTENANCE 27 10 4 QUALITY INSPECTION OF MEDICAL BEDS 28 10 5 CLEANING AND DIS...

Страница 4: ...ANSPORT AND STORAGE For transport the bed should be in its low position on a pallet and strapped and protected The wired control and supply lead should be attached to the bed base The head and footboa...

Страница 5: ...e side of the back rest or leg rest if this is not flat In the case of a prolonged more than 50 tilt bust semi sitting position it is recommended to vary the position of the person in bed every 2 hour...

Страница 6: ...n the designated room foreseeing an appropriate perimeter of use for the different functions variable height TR etc especially if the bed has a lifting pole or side rails Check that there is sufficien...

Страница 7: ...enetration Index 1st number decade Protection against solids 2nd number unit Protection against water intrusion 0 No protection No protection 1 Protected against solid bodies greater than 50 mm Protec...

Страница 8: ...ould be at least 30 Electrical fast transients EN 61000 4 4 2 kV for feeders 1 kV for input output lines 2 kV for feeders 1 kV for input output lines The quality of the main power supply must be the s...

Страница 9: ...an 3 V m Recommended separation distances between portable and mobile RF communications equipment and the medical bed see references in contents The medical bed see references in contents is intended...

Страница 10: ...8 cm minimum 1x Width 88 cm minimum with a high resilience foam of 27Kg m minimum Item Width 120 Width 140 Width 160 MEDIDOM II 2 P319 00 P320 00 P321 00 WINNEA P322 00 MEDIDOM II ELEGANT 2 P324 00 P3...

Страница 11: ...00 S A M ACTIV NA A622 00 Angled lifting pole 75 A623 00 Chromed angled lifting pole 75 A631 00 Bed base extension L140 cm NA A632 00 Bed base extension L160 cm NA A637 638 00 EPOXY side rail NA A639...

Страница 12: ...the bed is placed on the ground on its wheels It is imperative to raise the kickstand on the ship kit before transporting the bed 6 2 1 Residual risks DANGER RESULT REQUIREMENTS FOR THE USER Trapping...

Страница 13: ...l WINNCARE recommends the use of a patient lift or a dining table bases with less than 13 5 cm high Right back rest Bed base Right leg rest Headboard Left back rest Footboard Castors Angled lifting po...

Страница 14: ...ot end bed base TOTAL Bed 140 DUO C Version manual crank leg rest without boards 53 kg 31 5 kg 30 kg 114 5 kg Bed 140 DUO P Version electric folding leg rest without boards 57 5 kg 29 kg 38 kg 124 5 k...

Страница 15: ...ric folding legs with knees Control box Electrical connection A Remote control or right side if cap infrared option B Remote left or cable side if receiver option IR 1 Right back rest 2 Left electric...

Страница 16: ...nuous beeps at the push of a button the remote control It is necessary to reset the electrical assembly This manipulation is only possible with a V0 or V2 type remote control see No of version on the...

Страница 17: ...when the bed is empty to familiarise yourself with the bed functions Lifting and lowering of back rest Lifting and lowering of variable height Lifting and lowering of electric folding Simultaneous li...

Страница 18: ...t done the patient or another person who leans on the bed may fall 7 ASSEMBLING AND DISMANTLING 7 1 Assembling 1 Brake the wheels 2 Place the foot halt of the base on the base cross brace ensemble to...

Страница 19: ...smantling 1 Brake the wheels and place the bed base in low position 2 Unplug from the mains and disconnect the variable height jack and leg rest jack depending on option See stage instructions for dis...

Страница 20: ...k rest handle with one hand Push or lift to compensate the pressure exerted by the patient and unhook the clip by the connecting rod with the other hand The back rest jack will then pivot downwards d...

Страница 21: ...suscitation 8 2 Leg rest 8 2 1 Leg rest with manual crank C To lift lift the leg rest using the handle at the end To lower relieve the leg rest slightly or to its maximum with the hand to release the...

Страница 22: ...To positionning the side rail Insert amount of the side rail folded in the grip jaw on the head side of the bed to hear CLIC Tighten the Rondo screws of the grip jaws on the bed base To remove the sid...

Страница 23: ...th both hands until it locks Check that it is properly slot in To lower the side rail Raise the top side rail with one hand Press on the unlocking button with the other hand Support the rail as it low...

Страница 24: ...op side rail with both hands until it locks Check that it is properly slot in by trying to get it down without using the unlocking buttons 2 In high position Raise the top side rail with both hands to...

Страница 25: ...l Press on the two unlocking buttons with both hands Support the rail as it lowers 9 5 Angled lifting pole and IV stand In case of psycho behavioral problems of the patient it is necessary to fix the...

Страница 26: ...uring seals Put the motors back in place and put in the same direction as at the beginning Maximum patient weight Warning refer to user s manual Safe operational load Safety clip No maintenance should...

Страница 27: ...ication that the electrical leads are connected all along the metal jambs to prevent shearing of these leads when the variable height is being activated Verification that all of the electrical leads a...

Страница 28: ...controls displays lights and locking Bed specific side rails Space between head of bed and siderail 60 mm Space between the half siderails 60 ou 318 mm Condition mounting direction adaptation to bed...

Страница 29: ...ed according to the disinfectant manufacturer s instructions the drying time without human presence is often the same as the remanence time Refer to the instruction manual SANIVAP steam appliance with...

Страница 30: ...ntenance is 8 years for beds and 5 years for accessories and boards 10 7 Garanties All of our products carry a warranty against any manufacturing defect provided the normal conditions for use and main...

Страница 31: ...osition on wheels normal event Press down on the variable height to remove the wheels normal use The selected function of the bed does not work The function is locked on the remote control Unlock the...

Страница 32: ...e of the European Union and Council Directive 2006 66 EC of 6 September 2006 on batteries and accumulators and waste batteries and accumulators The complete product and or the easily detachable electr...

Отзывы: