6.6. Electrical system initialization
The control box can put itself in default (bad connection, cable fault, actuator fault ...).
In this case the movements are impossible and the control box makes short and discontinuous beeps at the
push of a button the remote control
It is necessary to reset the electrical assembly.
This manipulation is only possible with a V0 or V2 type remote control (see No. of version on the back of the
remote control).
Simultaneously press the two buttons of back rest and hold until the
end of a long and discontinuous beep.
Return the height variable actuator to the maximum