Guidance and manufacturer's declaration –
electromagnetic immunity to interference
The AIRcon Gen2 is intended for operation in an environment as specified below�
The customer or user of AIRcon Gen2 should ensure that it is operated in such an
environment�
Checks for
immunity to
interference
IEC 60601 –
test level
Conformity level
Electromagnetic
environment –
Guidelines
Electrostatic
discharge (ESD)
cf� IEC 61000-4-2
± 8kV
contact discharge
± 15kV
air discharge
± 8kV
contact discharge
± 15kV
air discharge
Floors should be
made of wood,
concrete or cera-
mic tile� If the floor
is covered with
synthetic mate-
rial, the relative
humidity must be
at least 30%�
Fast transient
electrical interfe-
rences / bursts cf�
IEC 61000-4-4
± 2kV for
mains lines
100kHz
repetition
frequency
± 2kV for
mains lines
100kHz
repetition
frequency
The quality of the
supply voltage
should be typical
for a business
or hospital
environment
Surges according
to IEC 61000-4-5
± 1kV differential
mode voltage
± 2kV common
mode voltage
± 1kV differential
mode voltage
± 2kV common
mode voltage
The quality of
the supply vol-
tage should be
typical for busi-
ness or hospital
environment