29
7.4.3 Warm-up time
During the warm-up time of 30 minutes,
certain alarm messages are suppressed�
Meanwhile, the icon
appears in the
lower right corner of the display� These
alarms indicate a deviation of an actual
temperature from that of the setpoint�
The reason for this lies in the fact that
the temperatures cannot be achieved
immediately after switching on�
7.4.4 Impact on the performance
There are no known adverse
effects on the performance of the
AIRcon Gen2 due to influences of
electrocauterization, electrosurgery,
defibrillation, X-rays, infrared rays,
transmitted switching pulses, magnetic
fields and radio frequency interferences�
7.4.5 Event list
By pressing and holding the lower right
and upper left buttons simultaneously,
the event list can be called up� It logs
the date, time and the alarm that
occurred� Up to 200 entries are can be
stored� If the memory is full, the last
entry is deleted and the new alarm is
added� The content of the event list
is retained even after turning off the
device�