WILAmed
AIRcon
28
KD:
Mute alarm�
Operation in alarm mode
LD:
Enter setup�
LD:
Change parameters
(setpoint temperatures and
expiratory heating capacity)�
7.4.1 Alarm muting
The audible alarm can be muted by
pressing the
button for 120 s and
can be terminated prematurely at any
time by pressing any key� Then the
alarm sounds again� If the cause of the
alarm is not resolved within 10 minutes,
the unit switches to "OFF" mode�
7.4.2 Alarm delaying
Alarm messages are also suppressed
for a certain time if the following
conditions are present:
{
{
After completion of treatment
interruption, the alarm delay is
activated for 3 minutes, because
during the treatment interruption
period no adjustment has been
made� Only the heating power was
reduced and the temperatures
cannot be immediately achieved
after switching to the normal mode�
{
{
After changing the humidification
level, the alarm delay is activated
for 3 minutes for the same reason�
7.4 Operation in alarm mode