GB
11258421.09
01/2012
GB/D
14
WIKA operating instructions temperature transmitter T32.xS
3. Specifications
Specifications
Models T32.xS, T32.xR
Permissible ambient temperature
range
{-50} -40 ... +85 °C
Climate class per IEC 654-1: 1993
Cx (-40 ... +85 °C, 5 ... 95 % relative air
humidity)
Maximum permissible humidity
■
Model T32.1S
per IEC 60068-2-38: 1974
■
Model T32.3S
per IEC 60068-2-30: 2005
Test max. temperature variation 65 °C
and -10 °C, relative humidity 93 % ±3 %
Test max. temperature 55 °C, relative
humidity 95 %
Vibration per IEC 60068-2-6: 2007
Test Fc: 10 ... 2000 Hz; 10 g,
Amplitude 0.75 mm
Shock per IEC 68-2-27: 1987
Test Ea: Acceleration Type I 30 g and
Type II 100 g
Salt mist per IEC 60068-2-52
Severity level 1
Free fall drop per IEC 68-2-52: 1996
Drop height 1500 mm
Electromagnetic compatibility
(EMC)
EMC directive 2004/108/EC, EN 61326
Emission (Group 1, Class B) and
immunity (industrial application)
{ } Items in curved brackets are options for additional price, not for rail mounting version T32.3S
The models T32.1R and T32.3R temperature transmitters
are available on request. The CE declaration of conformity
(ATEX/EMC) is valid without restrictions for T32.1R and
T32.3R (see Appendix 2 "EC Declaration of conformity").
For further specifications see WIKA data sheet TE 32.04 and the
order documentation.
For further important safety instructions for operation in
hazardous areas see chapter 10 "Notes for mounting and
operating in hazardous areas".
3. Specifications