24
WIKA operating instructions, model CPC2000
11155094.02 07/2012 GB/D/F/E
GB
6. Commissioning, operation
6.5.1.2 Unit
Here the unit, in which the low-pressure controller will display the measured value on the
screen, can be selected. Dependent upon the measuring range of the instrument, various
units are not available since they could not be displayed.
6.5.1.3 Steps
Here the step size for the STEP function, in x % of the set range, can be changed. Possi-
ble settings are from 1 … 50 %.
6.5.1.4 Mode
Here the operating mode of the controller can be set. The following modes can be
selected:
■
Measure (MEAS): The controller measures the pressure in the set pressure unit
■
Control (CTRL): A set point is automatically regulated.
■
Automatic (AUTO): A stored test routine will be carried out.
6.5.1.5 Language
Here the language in which the menu will be displayed can be set. There is the possibility
to choose between German, English, Spanish and French.
6.5.1.6 Settings
■
ZERO: Switching on and off of the automatic zero point adjustment following exit from
the menu. It is even possible to set at what intervals the instrument will automatically
perform a zero point adjustment.
■
RS232/USB: Selection of the appropriate serial interface
■
DISPLAY: Brightness setting
■
AUTO-MODE: Setting of the pressure range, the pressure unit and number of steps
the pressure range will be divided up into. The STEPS UP/STEPS DOWN can be
assigned different values.
■
INFO: Basic information on the instrument