19
WIKA operating instructions, model CPC2000
11155094.02 07/2012 GB/D/F/E
GB
6. Commissioning, operation
■
In the menu, go to the UNIT item with the
/
button, select the pressure unit and
confirm with SELECT.
■
Press the MENU button in order to exit the menu.
■
SETTINGS submenu:
■
Go to the SETTINGS menu item with the
/
button, confirm with SELECT.
■
In the submenu, go to the zero point adjustment item and confirm with SELECT.
■
Under the point, AUTO-ZERO, the automatic zero point adjustment for the MEAS
mode can be activated or deactivated.
Operation
On leaving the menu, the zero point of the internal sensor is adjusted, dependent upon
the setting. The instrument then begins to measure the subsequent pressures.
Using the ZERO button in MEAS mode, the instrument‘s zero point can be compensated
at any time.
CAUTION!
If the connected pressure should exceed the permissible highest loading
(125 %), the internal sensor will be closed off using a magnetic valve.
The display will then show ERROR. Using the SELECT button, the
instrument can be closed off and opened up freely.
Operating mode
Battery status
Measured value
Set point in % of the full scale value
Pressure unit
Full scale value