WIKA operating instructions, model CPC2000
11155094.02 07/2012 GB/D/F/E
GB
6. Commissioning, operation
6.3 Operating mode 'AUTO'
In AUTO mode, a pressure profile can be stored. This is especially useful when multiple
pressure sensors or pressure switches are to be calibrated with the same values. The
mode also enables a defined profile to be traversed several times in succession.
Setting
■
Press the MENU button. Confirm the menu item MODE with the SELECT button and
press the
/
button until AUTO appears. Confirm with SELECT.
■
For the AUTO mode, the submenu points RANGE, UNIT and STEPS are not relevant.
■
Go to the SETTINGS menu item with the
/
buttons and confirm with the
SELECT button. The submenu item zero point adjustment is not relevant.
■
In the submenu, go to the AUTO mode item with the UP/DOWN buttons and confirm
with SELECT.
■
RANGE: Pressure range that will be covered
■
UNIT: Pressure unit
■
STEPS: The number of steps which will be taken to increase and decrease the
pressure. These increments are calculated by the controller itself.
In the submenu point, CONFIGURATION 1, further settings can be defined for the AUTO