10.3 Selecting pulse length limiting
Debounces mechanical contacts via the filter frequency.
Parameter
Description
no
No evaluation of the pulse length.
2
2 Hz with a pulse-duty factor of 1:1 (minimum pulse length 250 ms)
5
5 Hz with a pulse-duty factor of 1:1 (minimum pulse length 100 ms)
10
10 Hz with a pulse-duty factor of 1:1 (minimum pulse length 50 ms)
20
20 Hz with a pulse-duty factor of 1:1 (minimum pulse length 25 ms)
50
50 Hz with a pulse-duty factor of 1:1 (minimum pulse length 10 ms)
100
100 Hz with a pulse-duty factor of 1:1 (minimum pulse length 5 ms)
500
500 Hz with a pulse-duty factor of 1:1 (minimum pulse length 1 ms)
10.4 Setting pulses per rotation
TYPE
FILt
[
▲
] [
▼
]
no
2
[
☼
]
[
▲
]
[
▼
]
[
☼
]
500
TYPE
PPt
[
▲
] [
▼
]
0001
…
9999
[
☼
]
[
☼
]