Operating instructions

Precision Pressure Transducer

Premium Version, Model CPT9000

EN

Precision Pressure Transducer, Model CPT9000

Содержание CPT9000

Страница 1: ...Operating instructions Precision Pressure Transducer Premium Version Model CPT9000 EN Precision Pressure Transducer Model CPT9000 ...

Страница 2: ...Operating instructions model CPT9000 12 2017 Mensor LP All rights reserved Mensor is a registered trademark of Mensor LP All other brand and product names are trademarks or registered trademarks of their respective companies Prior to starting any work read the operating instructions Keep for later use ...

Страница 3: ...3 7 Labeling Safety Marks 12 3 7 1 Product Label 12 3 7 2 Symbols 13 3 8 Warnings and Cautions 13 4 Transport Packaging and Storage 15 4 1 Transport 15 4 2 Packaging and storage 15 5 Installation 16 5 1 Mounting 16 5 1 1 Pressure Connections 16 5 1 2 Safety Relief Valves 16 5 1 3 Pressure Media 16 5 1 4 Electrical Connections 17 6 Operation 18 6 1 General Operation 18 6 1 1 Command and Query Forma...

Страница 4: ...d the Warranty 27 7 2 Recalibration 27 7 2 1 Calibration Services by Mensor or WIKA worldwide 27 7 2 2 Environment 28 7 2 3 Pressure Standards 28 7 2 4 Media 28 7 2 5 Setup 28 7 2 6 Calibration Communication Software 29 8 Dismounting Return and Disposal 30 8 1 Dismounting 30 8 2 Return 31 8 3 Disposal 31 9 Specifications 32 9 1 Measurement Specification 32 9 2 General Specification 33 9 3 Approval...

Страница 5: ...5 WIKA operating instructions premium pressure transducer model CPT9000 0020504001G 10 2019 EN EN ...

Страница 6: ... technical modifications Factory calibrations are carried out in accordance with international standards Further information Mensor Corporation Address 201 Barnes Dr San Marcos TX 68666 Internet address www mensor com Relevant data sheet CT 25 12 Application consultant Tel 1 512 396 4200 1 800 984 4200 USA only Fax 1 512 396 1820 E Mail sales mensor com techservices mensor com WIKA Alexander Wieng...

Страница 7: ...eration in industrial residential or commercial environments 1 3 Software License Agreement This product contains intellectual property i e software programs that are licensed for use by the end user customer hereinafter end user This is not a sale of such intellectual property The end user shall not copy disassemble or reverse compile the software program The software programs are provided to the...

Страница 8: ...8 WIKA operating instructions premium pressure transducer model CPT9000 0020504001G 10 2019 EN EN ...

Страница 9: ...as low as 10 H2O to as high as 15 015 psia Figure 2 1 2 1 Features Here is a short list of significant features designed into the CPT9000 Accuracy to 0 008 IS 33 Wide voltage power input range IP 67 rating for harsh environments All 316L SS wetted path on ranges 5 psi Command emulation for digital communications User selectable units Compact size 2 2 Scope of Delivery Precision Pressure Transducer...

Страница 10: ...ion 3 2 Intended Use The CPT9000 pressure transducer is designed to measure pressure with extreme accuracy as stand alone devices or as incorporated in OEM type applications It is characterized over a wide temperature range and is designed to be used in environments that are much more harsh than tolerable by previous generations of Mensor pressure transducers 3 3 Improper Use WARNING Injuries thro...

Страница 11: ...le injury and damage to equipment The activities described in these operating instructions may only be carried out by skilled personnel who have the qualifications described below Skilled personnel Skilled personnel authorized by the operator are understood to be personnel who based on their technical training knowledge of measurement and control technology and on their experience and knowledge of...

Страница 12: ... displayed in the work area regarding personal protective equipment The requisite personal protective equipment must be provided by the operating company Wear safety goggles Protect eyes from flying particles and liquid splashes Wear a protective helmet Protects the head from falling objects Wear safety shoes Protect feet from falling objects or objects lying around as well as against toxic or haz...

Страница 13: ...een trained in proper safety practices WARNING NOT EXPLOSION PROOF Installation of this instrument in an area requiring devices rated as intrinsically safe is not recommended WARNING POSSIBLE INJURY The tubing valves and other apparatus attached to the gauge must be adequate for the maximum pressure which will be applied otherwise physical injury to the operator or bystanders is possible CAUTION U...

Страница 14: ... the symbols on the packaging If the instrument is transported from a cold into a warm environment the formation of condensation may result in instrument malfunction Before putting it back into operation wait for the instrument temperature and the room temperature to equalize 4 2 Packaging and storage Permissible conditions at the place of storage Storage Temperature 40 85 C Humidity 5 95 relative...

Страница 15: ... Damage to the Instrument When working on open electrical circuits printed circuit boards there is a risk of damaging sensitive electronic components through electrostatic discharge The correct use of grounded working surfaces and personal armbands is required 5 1 Mounting The transducer may be mounted via its process connection or with any ring clamps along its body Having the sensor oriented in ...

Страница 16: ...5 Cable connections are External Connector Pinouts Pin Function Pigtail Color 1 9 18V Brown 2 RS 485 TX White 3 RS 485 RX Blue 4 RS 485 RX or RS 232 RX Black 5 Ground Gray 6 RS 485 TX or RS 232 TX Pink HOST DB9 COM Female GND PWR 12 VDC GRAY BLACK PINK 2 3 5 BROWN RS 232 HOST DB9 COM Female GND PWR 12 VDC GRAY BLACK PINK 2 3 5 BROWN RS 485 9 1 BLUE WHITE Figure 5 1 4 COM Connection Pin out and RS ...

Страница 17: ...y is programmable depending on what the user wants in the output string 6 1 1 1 RS 485 Command and Query Format On RS 485 versions of the CPT9000 the commands must be preceded by a an address or a In RS 485 the ADDRESS command sets the address in ranges from 0 to 9 and A to Z The global address character is the When data is required for a command a space only should delimit command and data fields...

Страница 18: ...ommand Set Case Return Value Valid data command Ready Invalid data command Invalid Data Invalid command Unknown Command Query Customer password protected command User Password Needed Table 6 1 1 3 Command set returns for given cases Also it is important to note that when sent data is determined to be valid the value will be stored in RAM Only after the SAVE command is sent is the data saved perman...

Страница 19: ...mands that are password protected are designated with Password Protected in the Query column below Global Commands Query Command Data Description ADDRESS ADDRESS sp C RS 485 only C Transducer address 0 9 A Z IDN ID N A Manu Model SN SW Version Manufacturer Model SN SW Version N A SAVE Saves Data to EEPROM TYPE N A C Pressure type 1 char FILTER FILTER sp n n Filter percentage 1 to 99 WINDOW WINDOW ...

Страница 20: ...T See Table 6 1 2 5 UNIT N A cccccccccc Unit text 10 char mmHg psi kPa CUST_UNIT CUST_UNIT sp n n nnnnnnnE nn Sets custom unit multiplication xx psi See Section 6 1 2 6 Calibration Related Commands CAL_DATE CAL_DATE sp yy mm dd Password Protected nn nn nn Calibration date SPAN CAL_SPAN sp n Password Protected n nnnnnnnE nn Customer span multipler 0 99 to 1 01 See Section 6 1 2 7 ZERO CAL_ZERO sp n...

Страница 21: ...Command Set Password Protection In the Sensor Command Set data that affects calibration is protected by a user settable 4 character password designated as Password Protected The default password is 0000 This is changed using the PWD_CHANGE command It should be changed once the user receives the device The factory password will override the user s password in the case that the user loses their pass...

Страница 22: ... NOT DETECTED The microcontroller bootloader was not detected 6 I2C TIMEOUT The microcontroller timed out trying to read or write using I2C 7 UART BUFFER OVERFLOW The UART buffer received more than 512 bytes of data 8 ERROR QUEUE IS FULL The error queue is now full This is the last error to fill the stack 9 OUT OF CAL The transducer exceeded calibration window 10 EEPROM RUNNING OUT OF SPACE The EE...

Страница 23: ...C 15 mbar 35 hPa 16 mmH2O 4C 36 Mpa 17 cmH2O 4C 37 mmH2O 20C 18 MH2O 4C 38 cmH2O 20C 19 mmHg 0C 39 mH2O 20C 20 cmHg 0C 99 CUST_UNIT Table 6 1 2 5 Available Unit Codes and Units 6 1 2 6 Customer Units The CUST_UNIT multiplier in conjunction with UNIT_INDEX 99 allows the user to set their own units The CUST_UNIT is a multiplier times psi to achieve the customer s unit value 6 1 2 7 Calibration Adjus...

Страница 24: ...put string 64 n 2 Address 1 Address added to all outputs 128 n 2 Table 6 1 2 10 Query Mode Mask Description 6 1 2 11 Burst Mode Output Setting BURST_MASK This sets the mask of the output parameters in burst output mode The transducer has the ability to continually output a limited set of binary data based on the selections set by the BURST_MASK variable The selected parameters will populate the ou...

Страница 25: ...MD_SET sp Y R Change CPT9000 into different command sets 1 is Mensor Legacy 0 is Sensor command set None DC XDC X sp DC sp Y Returns the date of calibration where Y is mmddyy Cust DC Y XDC sp Y R Loads the date of calibration where Y is mmddyy None DT XDT X sp DT sp NNN Returns the temperature a d DAC offset None FL XFL X sp FL sp NN Returns the filter it is the percentage of old reading added to ...

Страница 26: ...ultiplier None T XT X sp T sp C Returns pressure type None W XW X sp W sp NN Returns the filter window in FS Mens W NN XW sp NN R Sets the filter window in counts Readings inside this window are filtered None ZC XZC X sp ZC sp Y Returns customer zero correction value in current units Cust ZC X XZC sp Y R Sets customer zero correction value in current units Table 6 1 3 Mensor Legacy Command Set Mai...

Страница 27: ...ncement of discontinuation typically 8 to 10 years All repairs should be performed by Mensor due to the complexity of performing these repairs 7 2 Recalibration The CPT9000 automatically calculates the pressure reading for all effects of temperature hysteresis repeatability and non linearity within the compensated temperature range The process is referred to as dynamic compensation because each re...

Страница 28: ...a compatible with 316L SS is acceptable A reference height variation between the standard and the CPT9000 can cause errors A calculation should be made to compensate for this difference 7 2 5 Setup The schematic in Figure 7 2 5 shows a typical setup used to calibrate the CPT9000 Deadweight Tester or Primary Standard Power Supply Absolute Diaphragm Capacitance Gauge Transducer Under Test Vent Suppl...

Страница 29: ...ust the rich feature set including Changing filter settings Modifying custom identification strings Showing the CPT9000 s error stack Setting the tare pressure on the CPT9000 Modifying the output format of the sensor Changing the pressure units output It also has many features to adjust various calibration parameters after the appropriate password is sent Setting the calibration date Setting the c...

Страница 30: ...n plants and compressors there is a danger of physical injuries and damage to property and the environment Before storage of the dismounted instrument following use wash or clean it in order to protect persons and the environment from exposure to residual media Wear the requisite protective equipment see Chapter 3 6 Personal protective equipment Observe the information in the material safety data ...

Страница 31: ...truments with lithium ion rechargeable batteries or lithium metal batteries The included lithium ion rechargeable batteries or lithium metal batteries are subject to the requirements of the dangerous Goods Act For shipping special packaging and labelling requirements must be observed A dangerous goods expert must be consulted when preparing the item to be shipped Do not send any damaged or defecti...

Страница 32: ... psig 0 1 to 0 100 bar 0 0 36 to 0 15 psig 0 25 mbar to 0 1 bar Bi directional 3 15 145 to 15 1500 psi 1 10 to 1 100 bar 0 18 0 18 to 15 145 psi 12 5 12 5 mbar to 1 10 bar Absolute pressure 0 15 to 0 1515 psia 0 1 to 0 101 bar abs 0 5 to 0 15 psia 0 350 mbar abs to 0 1 bar abs AND 0 1515 to 0 15 015 psia 0 101 to 0 1001 bar Calibration interval 365 days Pressure units 39 and 1 user defined CPT9000...

Страница 33: ... wetted parts Voltage supply Power supply 9 to 18 VDC 12 VDC nominal Power consumption 26 mA at 12 VDC 5 Permissible ambient conditions Compensated temperature range 0 to 50 oC 32 to 155 oF Operating temperature range 40 to 85 oC 40 to 185 oF Storage temperature range 40 to 85 oC 40 to 185 oF Humidity 0 95 r h non condensing Operating altitude 3000 meters 10 000 feet Operating environment IP 67 Me...

Страница 34: ...108 9mm 4 29in 85 7mm 3 37in 18 9mm 74in 31 8mm 1 25in 6 PIN M8 CONNECTOR 1 16 BARBED REFERENCE PORT 1500 psi 1500 to 6000 psi 6000 to 15000 psi SAE J514 37 FLARE PORT 7 16 20 THREAD 117 6mm 4 63in 85 7mm 3 37in 18 8mm 74in 31 8mm 1 25in 6 PIN M8 CONNECTOR SEAL SCREW SAE J514 37 FLARE PORT 7 16 20 THREAD 122 9mm 4 84in 25 3mm 1 00in 31 8mm 1 25in 95 0mm 3 74in 6 PIN M8 CONNECTOR SEAL SCREW AUTOCLA...

Страница 35: ... the transducer have the following Mensor Part numbers Part Number Description 6000604023 Male NPT 6000604024 7 16 20 SAE 6000604025 BSPP 4250040006 Copper gasket washer Many other fittings are available consult the factory if you need assistance 10 2 Accessories Part Number Description 4050060021 Communication Cable 0020528001 Power Supply RS 232 communication cable 0020528002 Power supply and RS...

Страница 36: ...ander Wiegand SE Co KG Alexander Wiegand Straße 30 63911 Klingenberg Germany Tel 49 9372 132 0 Fax 49 9372 132 406 info wika de www wika de Mensor LP 201 Barnes Drive San Marcos TX 78666 USA Tel 1 512 3964200 15 Fax 1 512 3961820 E Mail sales mensor com www mensor com WIKA subsidiaries worldwide can be found online at www wika com ...