background image

257

eMc/eMi compliance

The UNI-DEX complies with the following EMC/EMI standards:

standard

test type

note

47 CFR  Part 

15, subpart C

RF emissions

USA Federal Communications 

Commission (FCC) requirements to 

intentional radiators.

EN 300 

330-2 V1.5.1

RF emissions 

incl. Spurious 

emission

EMC and radio spectrum matters 

for Short Range Devices in the 

frequency range 9 kHz – 25 MHz

IEC 60601-1-

2:2007

EMC emission

Immunity, RF 

and ESD

Medical electrical equipment.

General requirements for basic safety 

and essential performance. 

Electromagnetic compatibility.

EN 301 489-3 

V1.4.1

Immunity, RF 

and ESD

Standard for Low Power Transmitters 

in the frequency range  

9 kHz – 40 GHz

9 514 0259 099_booklet.indb   257

14-11-2013   13:22:42

Содержание UNI-DEX

Страница 1: ...UNI Dex 9 514 0259 099 01 omslag indd 7 14 11 2013 13 00 09...

Страница 2: ...1 6 1 5 1 4 1 2 1 1 1 3 1 7 3 2 3 1 3 3 4 2 4 1 5 1 5 2 5 3 5 4 1 4 5 2 3 1 2 3 2 2 2 1 2 3...

Страница 3: ...e jack which is common in newer mobile phones If your mobile phone or audio source uses a different configuration it may not work with this product Accessories See illustration 2 2 1 Charger 2 2 Plug...

Страница 4: ...cation out of reach of children and pets Never try to open or repair the device yourself This may only be done by authorized personnel Troubleshooting Problem Possible cause Solution The device does n...

Страница 5: ...ly connect a charger that is compatible with the device The charger must have an output rating of 5VDC and a micro USB connector The charger input rating voltage and wall plug must be compatible with...

Страница 6: ...least 15 cm 6 inches away from the implant can be specified as Autophone magnet hearing instrument case magnet in a tool etc Hereby Widex A S declares that this UNI DEX is in compliance with the esse...

Страница 7: ...ture and or The product must be kept away from rain symbol title Description lower limit of temperature The lowest temperature to which the product can be safely exposed upper limit of temperature The...

Страница 8: ...ty with the requirements set out in European CE marking directives symbol title Description alert The product is identified by R TTE Directive 1999 5 EC as an equipment Class 2 product with some restr...

Страница 9: ...o Modulation FSK Radiated Output Power 85 2 dBm 3 02pw ERP effective radiated power Magnetic field strength 26 9 dB A m at 10 m distance Duty Cycle Up to 100 averaged over 1 hour of operation Simplex...

Страница 10: ...s internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The UNI DEX is suitable for use in...

Страница 11: ...harge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at leas...

Страница 12: ...5 dip in UT for 5 s Not applicable Not applicable Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels characteristic of a typical loca...

Страница 13: ...l Compli ance level Electromagnetic environment guidance Portable and mobile RF communications equipment should be used no closer to any part of the UNI DEX including cables than the recommended separ...

Страница 14: ...m fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked...

Страница 15: ...MHz field strengths should be less than 3 V m Recommended separation distances Recommended separation distances between portable and mobile RF communication equipment and the UNI DEX The UNI DEX is i...

Страница 16: ...ter in watts W according to the transmitter manu facturer NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propa...

Страница 17: ...s incl Spurious emission EMC and radio spectrum matters for Short Range Devices in the frequency range 9 kHz 25 MHz IEC 60601 1 2 2007 EMC emission Immunity RF and ESD Medical electrical equipment Gen...

Страница 18: ...nce with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does caus...

Страница 19: ...ce including interfe rence that may cause undesired operation of the device Conform ment la r glementation d Industrie Canada le pr sent metteur radio peut fonctionner avec une antenne d un type et d...

Страница 20: ...WIDEX A S Nymoellevej 6 3540 Lynge Denmark www widex com 5qr0w6 ll g Manual no 9 514 0259 099 01 9 514 0259 099 01 omslag indd 6 14 11 2013 13 00 08...

Отзывы: