254
Radiated RF
IEC
61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
d = 1.2
√
P
80 MHz to 800 MHz
d = 2.3
√
P
800 MHz to 2.5 GHz
Where
P
is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey
a
, should be less than
the compliance level in each
frequency range
b
.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propa-
gation is affected by absorption and reflection from structures, objects and peo-
ple.