250
guidance and manufacturer’s declaration
Electromagnetic emissions
The UNI-DEX is intended for use in the electromagnetic environment speci-
fied below. The customer or the user of a UNI-DEX should assure that it is
used in such an environment.
emissions
test
compliance
electromagnetic environment
- guidance
RF
emissions
CISPR 11
Group 1
The UNI-DEX uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF
emissions
CISPR 11
Class B
The UNI-DEX is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic
emissions
IEC
61000-3-2
Not
applicable *)
Voltage
fluctua-
tions/ flicker
emissions
IEC
61000-3-3
Not
applicable *)
*) Battery powered equipment
9 514 0259 099_booklet.indb 250