WHEATON UniSpense
®
PRO Peristaltic Pump
WHEATON | 800.225.1437 (U.S. & Canada Only) | 856.825.1100 | www.wheatonsci.com| P/N 50099073 Rev. 2/2/10 Lit. No.8734
20
14.3 Fuse Replacement
DANGER! BE CERTAIN THE UNIT IS DISCONNECTED FROM THE AC POWER
SOURCE.
1. Disconnect the cord-set from the AC power source.
2. Remove the fuse drawer by using a small flat screwdriver to lift
the tab.
3. Replace with new fuses:
108 - 120 vac unit use (2) 5x20mm, 250V, 2.0 AT fuses,
be sure the correct cord set is used.
220-240 vac unit use (2) 5x20 mm 240V 1.0 AT fuses
, be sure the correct cord set is used.
15.0 Illustrations and Diagrams