7. Troubleshooting

Please contact our technical support team on 1800 358 101. 

This centrifuge has a self-diagnostic function. If an issue occurs, an error or  

warning code will become visible on the display screen. The user can determine the 

malfunction with the below error codes. 

Error Code

Cause

Corrective Steps

No display

No main power connection

Check the power and the mains 

cable at each end.

Power failure

Check the mains fuse of the lab.

Improper connection

Connect the adaptor properly.

 

Lid incorrectly closed

Close the lid correctly.

Error with the lid’s closing and 

opening mechanism

Contact the technical support 

team on 1800 358 101.

Err 55

Rotor loaded asymmetrically

Load the rotor symmetrically 

and restart the centrifuge.

Lid cannot 

be opened

Spinning rotor

Wait for the rotor to come to a 

stop.

Power failure

The emergency lid release 

safety feature will trigger once 

the rotor stops.

Shakes 

during 

acceleration 

and irregular 

operating 

feedback

Rotor loaded asymmetrically

Load the rotor symmetrically 

and restart the centrifuge.

Broken tube

Replace the broken tube.

Damage to the rotor  

or the motor

Contact the technical support 

team on 1800 358 101.

Damaged rotor

Remove and change rotor.

Display error

Loose display connection

Contact the technical support 

team on 1800 358 101.

Err 1

Damaged or jammed latch

Contact the technical support 

team on 1800 358 101.

Err 52

Motor is jammed or operating 

voltage is incorrect

Turn the centrifuge OFF. Check 

the rotor fitment or apply the 

correct operating voltage.

Power 

tripping

Cable incorrectly fitted

Remove the cable and connect 

correctly.

Содержание 4500RPM Max

Страница 1: ...1 PRODUCT MANUAL Multipurpose Lab Centrifuge 4500RPM Max Code 665 113...

Страница 2: ...ion the unit can halt and resume operations without disturbing fragile samples To protect the samples further the centrifuge s intelligently designed airflow system ensures that the temperature of the...

Страница 3: ...mum Rotor Capacity 4 x 100ml Maximum Speed RPM RCF g 4500 RPM g varies per rotor Speed Setting Range 500 to 4500 RPM Step of 100 RPM Speed Accuracy 100 RPM Time Setting Range 1 to 99 minutes and infin...

Страница 4: ...or the rotor chamber the centrifuge must be cleaned carefully before resuming operations The centrifuge may be used for the specified applications only It must not be operated in a hazardous or flamm...

Страница 5: ...tamination caused by aggressive agents the rotor must be cleaned immediately with a natural cleaning liquid Pay particular attention to the bores of the tubes If any damage is seen contact the service...

Страница 6: ...ed as a lid open function of the centrifuge motor is not running 3 RPM RCF Single press to display the RPM and for the RPM LED to blink Long press the button to toggle the display between RPM and RCF...

Страница 7: ...N the power switch located on the right of the unit ATTENTION Maintain a gap of three seconds between turning the power switch OFF and then turning it ON again Do not turn ON and OFF again immediately...

Страница 8: ...ters Speed Time ACC DCC as described above To select the programs press the increment and decrement buttons All the parameters in the program will save automatically after the display stops blinking T...

Страница 9: ...15ml 4500 3485 L30 140 Rotor 25 32 x 10ml 4500 3164 L30 124 Rotor 26 2 x Microplates 4500 2377 Fixed Angle Rotors Rotor Code Rotor Max Volume Max RPM Max RCF L30 136 Rotor 7 4 x 100ml 4500 2559 L30 1...

Страница 10: ...osing and opening mechanism Contact the technical support team on 1800 358 101 Err 55 Rotor loaded asymmetrically Load the rotor symmetrically and restart the centrifuge Lid cannot be opened Spinning...

Страница 11: ...ed States Canada Error Code Cause Corrective Steps Last run memory not displayed Turning the centrifuge ON immediately after turning it OFF Maintain a three second gap between turning the centrifuge O...

Страница 12: ...Westlab in every discovery...

Отзывы: