Note: the regression is computed on the points included within the left and right
cursors. All the other points are ignored by the calculation.
The selection of the flux curve interval where to compute the flux is critical and
a wrong selection can cause an error on the estimation of the flux. In the
following pages some examples of the correct interval selection are shown.
To change the interval move the left and right cursors to the desired position.
To move the cursors you have to touch the screen close to the limit you want
to move and drag it to the correct position. The software will move the limit that
is closer to the point you touch on the screen.
Once having selected the interval FluxManager will compute the regression and
the results will be shown as text and as best fit line.
Each flux curve (e.g. CH
4
, H
2
S, CO
2
) has a distinct regression interval, then after
the computation of the regression of one flux curve you have to select, using
the
NEXT
button, the other gas flux curve and select the appriopriate regression
interval.
Left cursor
Right cursor
Flux data
Regression line