5 Components
W
EST
Syst
em
s
PAG
E
46
w
w
w
.w
es
tsy
stems
.co
m
W
EST
Syst
em
s
The charger is started by connecting the battery pack to the charger and then
connecting the charger to the mains. The LED will be yellow before the fast
charge starts and the LED changes to orange. When the batteries are fully
charged and the voltage drops because of the -dV signal from the batteries, the
charger will go into a top-off charge mode before it goes over to trickle charge
mode. During top-off charge the LED will be green with a short intermittent
yellow light. When the top-off charge is completed, the charger will go into
trickle charge mode and the LED will be green. The charge current is now
reduced to a safe level, which allows the charger to stay connected to the NiCd
batteries without damaging the batteries. NiMH batteries are not as well suited
for trickle charge, and some battery manufacturers recommend that trickle
charge
does not exceed 24 hours. If in doubt; contact the battery manufacturer
for details. If the safety timer runs out before
–
dV is detected, the charger will
go directly to trickle charge mode (no top-off charge) and LED will be
continuously green. lf the battery voltage is far below normal, the charger will
cut the fast charge current and go to trickle charge mode. The LED will then
indicate
”error” by flickering green and orange light.
lf the mains input voltage
is turned off, the charger will reset. When the mains input voltage is turned on
again a new charge cycle will start. lf new batteries are to be connected, the
charger must idle for approx 15 sec. to make sure all parameters in the
microprocessor have been reset. When the charger has been reset the LED
changes to yellow, and a new charge cycle can begin.
LED indications
LED
MODE
YELLOW
Battery not connected
YELLOW
Battery initialisation and analysis
ORANGE
Fast charge
GREEN with intermittent
YELLOW flash
Top-Off Charge
GREEN
Trickle Charge
Alternating ORANGE-
GREEN
ERROR