Q6: What kind of security is available in recorder?
A6: There are two kinds of security. The normal security offers a common password. The high
security with FDA 21 CFR part 11 compliance is available from version 2.2 on both
recorder and PC software Observer. It supports maximum 30 users with different
passwords. The administrator has the right to define supervisor and operators. The
supervisor can change configuration, save and clear data but not operators. Operators
can only view the historical data, events and status.
Q7: What is Boot Rom? Anything related to Math option? Is it possible to upgrade it into new
version via internet or CF card?
A7: Boot ROM is a chip available in recorder, which contains boot loader information. The
Boot ROM BT181 has no option. The Boot ROM BT182 has options of Math, Counter,
Totalizer & FDA 21 CFR part 11 compliance. It is required to remove BT181 away and
install BT182 on the socket when those options are needed. It is not possible to upgrade
it into new version via internet or CF card.
Q8: How does it upgrade into the new version of recorder system and PC software?
A8: If necessary, contact your supplier about the latest updates. The supplier is able to
supply the latest version of recorder system and PC software via email or FTP ( file
transfer protocol ) for easy download through internet. Follow the procedure to update it
into the new version of recorder system. Install the new version of Observer software. It
shall automatically uninstall the old one.
Q9: If it needs to upgrade the recorder system, what do we have to take care?
A9: Any recorder system before V 1.20, it is mandatory to update first to V 1.20 and then to the
latest version i.e. it needs to upgrade twice. Direct update to the latest version could
damage the DOC.
Q10: What are the differences between Observer I & Observer II PC software?
A10: Observer I is supplied for free. Observer II is supplied at additional cost. Observer I is
used for recorder configuration from PC and also it is used for reading historical data in
PC at set time period of 1 to 24 hours. Observer II is used for real time data monitoring in
PC along with configuration and historical data view.
Q11: How many AI, AO, DI, DO cards, and 24VDC auxiliary power card can be inserted into
the recorder?
A11: Maximum 6 cards in any combination can be inserted into rear Slots.
Q12: What are those larger capacities of compact flash cards supported by the recorder and
recommended manufacturer?
Содержание WH18/06
Страница 1: ...User Manual Paperless Recorder WH18 06 15th Edition 03 2008 ...
Страница 17: ...17 Figure 2 6 Figure 2 7 Figure 2 8 ...
Страница 19: ...19 Figure 2 10 Figure 2 11 ...
Страница 25: ...25 Digital input card DI181 Figure 2 21 Analog output card AO183I AO183V Figure 2 22 ...
Страница 26: ...26 24 VDC auxiliary power supply card AP181 Figure 2 23 2 5 RS 232 RS 422 RS 485 wiring Figure 2 24 ...
Страница 27: ...27 Figure 2 25 Figure 2 26 ...
Страница 70: ...70 Make sure that RS 232 communication setting at the recorder and PC are equal ...
Страница 87: ...87 7 Trouble shooting ...
Страница 88: ...88 ...
Страница 89: ...89 ...
Страница 90: ...90 ...