7 Measurement
7.1 Ordinary measurement
7.1.1 Distance measurement
1.
First freeze the image.
2.
Press “
+/9
”or the left button of mouse to display the measurement cursor.
3.
Move the cursor to measurement start point.
4.
Press “
” or the right button of mouse, and move the cursor to measurement end point. (press
“
” or the right button of mouse to switch the start and end cursor)
5.
Press middle button of mouse to end the measurement.
Repeat steps 2-5 to measure continuously. Up to 4 groups of measurement is allowed. If you want to
continue the measurement after 4 groups, the original data would be overwritten. The current measure
result would be displayed on the right of screen. The distance rate is listed as “D1/D2, D3/D4”. Press “
B/0
”
to clear the measurement mark and measurement result.
7.1.2 perimeter/area measurement
Free measurement
1.
First clean the origin date on the screen.
2.
Press “
FUNC/4
”→“7”in turn to display following dialog box:
3.
Press “1” to choose free method, and the measure cursor is displayed on the screen.
4.
Move the cursor to measurement start point.
5.
Press “
” or right button of mouse, move the cursor to end point through the edge of measure
areas to form a closed path.
6.
Press “
” or right button of mouse to complete the measurement. If the curve is unclosed, the
PLEASE ENTER:
1.FREEHAND 2.ELLIPSE
Distance measurement chart