Calibration
To calibrate, proceed as follows:
THE UNIT MUST BE CALIBRATED WITH THE FIRST PULMONARY
FUNCTION TEST OF THE DAY, AFTER EVERY SIGNIFICANT
TEMPERATURE CHANGE OR AFTER CHANGING THE SENSOR.
Select the Calibration menu option (ALT key then key '8').
The following screen is displayed:
ALT
then
8
25
>
Last calib.
:
25.03.99
BTPS Factor
:
1.090
Calib. Factor
:
1.002
Temperature
:
21
o
C
Measured volume
:
0.00 l
Syringe Vol.
:
6.00 l
Calibration
To calibrate, proceed as follows:
THE UNIT MUST BE CALIBRATED WITH THE FIRST PULMONARY
FUNCTION TEST OF THE DAY, AFTER EVERY SIGNIFICANT
TEMPERATURE CHANGE OR AFTER CHANGING THE SENSOR.
Select the Calibration menu option (ALT key then key '8').
The following screen is displayed:
ALT
then
8
25
>
Last calib.
:
25.03.99
BTPS Factor
:
1.090
Calib. Factor
:
1.002
Temperature
:
21
o
C
Measured volume
:
0.00 l
Syringe Vol.
:
6.00 l