SYSTEM SETTINGS
Page | 75
WARNING:
When the 40 Hz filter is used, the frequency response requirement for diagnostic ECG
equipment cannot be met. The 40 Hz filter significantly reduces high-frequency components of the ECG
and pacemaker spike amplitudes, and is recommended only if high-frequency noise cannot be reduced by
proper procedures.
NOTE: A filter setting lower than 150 Hz will reduce the visibility of fast transients in the ECG like
pacemaker spikes or fast notches. For pediatric ECGs a 300 Hz setting is recommended. Filter settings
apply only to displayed and printed data. Data is stored in unfiltered format.
NOTE: The High Pass filter (or base line filter), indicated by the number "0.05" cannot be changed.
RScribe automatically implements a high-performance base line filter that does not distort the ECG
waveform. High Pass filters that do distort the ECG waveform are not available.
AC Filter
To change default ECG AC filter:
Select
AC Filter
Select from: None, 50 Hz, or 60 Hz
NOTE: RScribe removes 60 Hz or 50 Hz interference. The setting you select depends on the line
frequency in your country. For example, use the 60 Hz setting in the U.S. If the setting is correct but you
still see mains interference, check the electrode connections, mains interference sources like transformers
or motors close to the patient, and the connection to the safety ground of the computer. Try operating from
battery power if needed.
Cabrera
To change default ECG to Cabrera:
Left mouse click on the
Cabrera
icon
Choices: On, Off
NOTE
:
Using the lock indicator to the right of this selection can be enabled to hide this choice from the
technician, permitting only unlocked choices.
Median Zoom
Lead
To change default median ECG lead format
display
:
Select
Lead
Choices: I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 , All leads (all 12 leads
superimposed)
Speed
To change default speed
on the display
:
Select
Speed
Choices: 100 mm/s, 200 mm/s
Gain
To change default ECG gain
:
Select
Gain
Choices: 10, 20, 40, 80 mm/Mv