Service Manual
Troubleshooting and repair
19
Service chart
This chart shows which procedure to use for specific service operations on monitors not
covered by warranty.
Monitor
Caution
Do not service a monitor under warranty. Any service not performed at
a Welch Allyn service center voids the product warranty.
Note
Do not attempt to replace any component, including the NIBP pressure
transducers, soldered onto the main board. All service of the main board must be
performed at the Welch Allyn factory service center.
Service
Action
Replace the NIBP air filter.
Replace the battery pack.
“Removing the lithium ion battery assembly”
Replace the connector panel
1.
“Removing the patient connector panel from the
2.
Replace the LCD assembly
1.
“Disassembling the top chassis”
2.
3.
Replace the top case assembly (with button board and window)
1.
“Disassembling the top chassis”
2.
Replace the bottom case assy (with power switch, power flex
cable, and connector panel)
1.
“Removing the main board from the bottom
2.
Replace the NIBP pump, bleed and dump valves, check valve, or
internal tubing
1.
“Removing the NIBP assembly from the main
2.
3.
Update the software
Send the monitor to a Welch Allyn service center.
(See
Replace the main board
Send the monitor to a Welch Allyn service center.
(See
Содержание Propaq LT 802LT0N
Страница 1: ...Propaq LT Vital Signs Monitor Service Manual Software version 1 5X or later ...
Страница 4: ...iv Contents Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 8: ...4 Safety summary Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 28: ...24 Troubleshooting and repair Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 74: ...70 Functional verification Welch Allyn Propaq LT Vital Signs Monitor ...
Страница 78: ...74 35 ml test volume Welch Allyn Propaq LT Vital Signs Monitor ...