Disassembly and reassembly
45
Removing the patient connector panel from the bottom chassis
1.
“Removing the lithium ion battery assembly”
2.
“Separating the top and bottom chassis halves”
3.
“Removing the main board from the bottom chassis assembly”
4.
Remove the clear plastic cover from the ferrite beads by removing the two screws
(one is under a strip of foam tape) securing it to the chassis.
5.
Pry the ferrite beads from the adhesive on the bottom case.
6.
The connector assembly is held in place by the two snap-in clips and by an adhesive
gasket. Remove the assembly by pushing in the clips while applying upward pressure
on the assembly to break the grip of the gasket.
REASSEMBLY:
Caution
The ferrite beads are extremely brittle and can chip or crack if
mishandled.
Caution
The clips are fragile. Excessive bending will break them.
Reassembly: 4 inch lbs
screwdriver tip
clip detail