43
TROUBLESHOOTING
The following table describes error and lead fail messages that are displayed on the H12+ recorder during
patient hookup and recording.
Table of Messages
Message
Solution
‘LOW BATTERY’
Replace existing battery with a fully charged battery.
‘CHECKSUM ERROR’
If checksum does not match the stored checksum at power up. Contact Welch Allyn
Technical Support Group.
‘CARD ERROR’
SD card error. SD card product capabilities identification file is missing. Contact
Welch Allyn Technical Support Group.
‘No card’
SD card not detected in the card slot. Install a formatted SD card.
‘CARD NOT Hillrom
FORMATTED’
SD not properly formatted. Contact Welch Allyn Technical Support Group.
‘CARD NOT ERASED’
Data detected on SD card. Erase the card using HScribe system software.
‘RATE_DUR.TXT error’
Rate/Duration file is not properly formatted or is set to exceed the limits. Press any
key to continue the powerup sequence.
‘PATINFO4.TXT error”
Patient information file is not properly formatted. Reformat the SD card at HScribe.
‘RA’
RA fail. Check if the lead wire is off or the electrode needs to be replaced.
‘RL’
RL fail. Check if the lead wire is off or the electrode needs to be replaced.
‘LA’
LA fail. Check if the lead wire is off or the electrode needs to be replaced.
‘LL’
LL fail. Check if the lead wire is off or the electrode needs to be replaced.
‘V1’
V1 fail. Check if the lead wire is off or the electrode needs to be replaced.
‘V2’
V2 fail. Check if the lead wire is off or the electrode needs to be replaced.
‘V3’
V3 fail. Check if the lead wire is off or the electrode needs to be replaced.
‘V4’
V4 fail. Check if the lead wire is off or the electrode needs to be replaced.
‘V5’
V5 fail. Check if the lead wire is off or the electrode needs to be replaced.
‘V6’
V6 fail. Check if the lead wire is off or the electrode needs to be replaced.
Any combination of
‘RA, LA, RL, LL, V1, V2,
V3, V4, V5, V6’
More than one lead is in fail. Check the lead wires and electrodes.
Содержание Hillrom H12+
Страница 6: ...SERVICE AND SPARE PARTS 2...
Страница 8: ...NOTICES 4...
Страница 10: ...WARRANTY INFORMATION 6...
Страница 16: ...EQUIPMENT SYMBOLS AND MARKINGS 12 Global Trade Item Number Battery orientation and size Medical Device...
Страница 18: ...GENERAL CARE 14...
Страница 27: ...INTRODUCTION 23 H12 Recorder in Carrying Case...
Страница 40: ...CONFIGURING THE RECORDER 36...
Страница 46: ...MAINTENANCE 42...
Страница 48: ...TROUBLESHOOTING 44...