PATIENT PREPARATION
28
NOTE AND CAUTION: Placement of the Left Leg (LL) electrode in the original Mason-Likar
position increases the similarity of the acquired ECG with a standard 12-lead ECG and is therefore
recommended; however, clothing may interfere with this position and increase the amount of artifact. The
modified position may decrease the sensitivity of inferior ECG leads and cause axis shift with respect to
the standard 12-lead ECG. Accurate skin preparation and suitable clothing are the most important factors
in excessive artifact prevention.
Limb Electrode
Precordial Electrode
AAMI
IEC
Placement
AAMI
IEC
Placement
RA
R
On or below the right
clavicle as shown
V1
C1
Fourth intercostal space at
the right sternal border
LA
L
On or below the left clavicle
as shown
V2
C2
Fourth intercostal space at
the left sternal border
RL
N
Reference or ground lead,
should be placed in a stable
location of the body
V3
C3
Midway between V2 and V4
LL
F
Lower-left side of body in a
stable location, as close to
the hip as possible
V4
C4
Fifth intercostal space at the
left of the midclavicular line
V5
C5
Anterior axillary line on the
same horizontal level as V4
V6
C6
Mid-axillary line on the same
horizontal level as V4 and V5
Содержание Hillrom H12+
Страница 6: ...SERVICE AND SPARE PARTS 2...
Страница 8: ...NOTICES 4...
Страница 10: ...WARRANTY INFORMATION 6...
Страница 16: ...EQUIPMENT SYMBOLS AND MARKINGS 12 Global Trade Item Number Battery orientation and size Medical Device...
Страница 18: ...GENERAL CARE 14...
Страница 27: ...INTRODUCTION 23 H12 Recorder in Carrying Case...
Страница 40: ...CONFIGURING THE RECORDER 36...
Страница 46: ...MAINTENANCE 42...
Страница 48: ...TROUBLESHOOTING 44...