Sensor Accuracy Guide
Accuracy specifications are based on controlled hypoxia studies with healthy,
non-smoking adult volunteers over the specified saturation SpO
2
range. Pulse oximeter
SpO
2
readings were compared to SaO
2
values of drawn blood samples measured by
hemoximetry. All accuracies are expressed as + “X” digits. This variation equals plus one
standard deviation (+ 1 SD), which encompasses 68% of the population.
Nellcor Patents
Covered by one or more of the following US Patents: 4,802,486; 4,869,254; 4,928,692;
4,934,372; 5,078,136; 5,351,685; 5,485,847; 5,533,507; 5,577,500; 5,803,910; 5,853,364;
5,865,736; 6,083,172; 6,708,049; Re. 35,122 and foreign equivalents.
Pulse Rate Range
25 to 240 bpm
Pulse Rate Accuracy
±3 bpm
±3 bpm (low perfusion)
Table 15. OxiMax Sensor Models, Single Patient Use
Sensor Models
SpO
2
Range
70% to 100%
MAX-AI
± 2
MAX-PI
± 2
MAX-II
± 2
MAX-RI
1
± 3.5
1
The accuracy specification has been determined between saturations of 80% to 100%.
Table 16. OxiCliq Sensor Models, Single Patient Use
Sensor Models
SpO
2
Range
70% to 100%
OXICLIQ-PI
± 2.5
Table 17. Reusable Sensor Models
Sensor Models
SpO
2
Range
70% to 100%
D-YS (Infant to Adult)
± 3
D-YS and D-YSE
± 3.5
D-YS and D-YSPD
± 3.5
DS-100A
± 3
OXI-A/N (Adult)
Adult: ± 3
OXI-P/I (Pediatric/infant)
± 3
Содержание 4200B
Страница 3: ...Welch Allyn Spot Vital Signs Directions for Use...
Страница 6: ...iv Welch Allyn Spot Vital Signs...
Страница 22: ...14 Welch Allyn Spot Vital Signs...
Страница 28: ...20 Internal Configuration Welch Allyn Spot Vital Signs...
Страница 36: ...28 Welch Allyn Spot Vital Signs...
Страница 42: ...34 Welch Allyn Spot Vital Signs...
Страница 44: ...36 Welch Allyn Spot Vital Signs...
Страница 50: ...42 Welch Allyn Spot Vital Signs...
Страница 60: ...52 Welch Allyn Spot Vital Signs...
Страница 70: ...62 Welch Allyn Spot Vital Signs...
Страница 72: ...64 Warranty Welch Allyn Spot Vital Signs...
Страница 73: ......
Страница 74: ...Reorder No 4200 87E Material No 706272 Ver C...