![Welcare WET100 Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/welcare/wet100/wet100_instruction-manual-and-warranty-information_953909019.webp)
Instruction Manual & Warranty Information
19.
Troubleshooting
- continued
Error
Code
Problem
Solution
Er 6
In object temperature
mode, measurement
is being taken with
probe cover on the
device.
Remove probe cover
in object temperature
mode.
Hi
In Ear mode:
Temperature taken is
higher than 43.0°C
(109.4°F).
In Object mode:
Temperature taken
is higher than 100°C
(212°F).
Read Temperature
Taking on page 8
thoroughly, then check
the integrity of the
probe cover and take
a new temperature
measurement.
Lo
In Ear mode:
Temperature taken
is lower than 32.0°C
(89.6°F).
In Object mode:
Temperature taken
is lower than 0°C
(32°F).
The thermometer’s
battery is low.
Replace with two
new alkaline batteries
size AA.
If the problem is not solved by the troubleshooting guide or
you have any queries, please contact our service centre:
In Australia
- 1800 807 464
In New Zealand
- 0800 523 583
DO NOT
attempt to open or repair the product yourself