16.

 

Welcare  Ear Thermometer

Important Safety Information

 

- continued

Replace the Battery

Replace battery when 

 appears in the bottom of LCD 

display. 

(figure C.)

•  Open the battery cover using a tool such as a pen or 

blunt knife.

•  Slide the battery cover down as shown below. (figure D.) 
•  Remove the battery and install 2 new AA alkaline 

batteries as shown below. (figure E.) 

•  Clip the battery cover back on.

NOTE:

 

To protect the environment, discard the used 

batteries in accordance with local regulations.

Replacement Accessories

Replacement accessories can be purchased online or by 
contacting us directly:

Australia:

 1800 807 464 www.welcare.com.au

New Zealand:

 0800 523 583 www.jadavey.co.nz

Part Code

Part Name

WET01-60

Welcare Ear Thermometer Probes 
(60PK)

(figure E.)

(figure D.)

(figure C.)

Welcare_WET01_EarTherm_IM.indd   16

Welcare_WET01_EarTherm_IM.indd   16

3/12/21   2:21 pm

3/12/21   2:21 pm

Содержание WET100

Страница 1: ... Carefully Before Use Model WET100 Instruction Manual Warranty Information EAR AND OBJECT TEMPERATURE DETECTION SECOND READING Welcare_WET01_EarTherm_IM indd 1 Welcare_WET01_EarTherm_IM indd 1 3 12 21 2 21 3 12 21 2 21 ...

Страница 2: ...care Ear Thermometer Please keep this instruction manual in a safe place where it is readily available Note The illustrations in this manual may differ slightly from the product itself are_WET01_EarTherm_IM indd 2 are_WET01_EarTherm_IM indd 2 3 12 21 2 21 pm 3 12 21 2 21 pm ...

Страница 3: ...d ON or OFF 11 Change to Ear Mode or Object Mode 11 Reset Parameter Format 11 Probe Cover Detection and Eject Function 12 Measuring Ear Temperature 13 Fever Alarm 14 Measuring Object Temperature 14 Memory Function 15 Replace the Battery 16 Replacement Accessories 16 Cleaning and Maintenance 17 Storage 18 Troubleshooting 18 Specifications 20 Product Warranty 22 Warranty Form 25 Disposal 28 Welcare_...

Страница 4: ...meter Probe Cover for hygiene purposes DO NOT forcibly insert the probe into the ear If you feel any discomfort such as a pain during the measurement stop using the product immediately DO NOT use this product if suffering from an ear disease such as otitis externa or otitis media It may worsen the condition DO NOT use this product when the external auditory canal is wet such as after swimming or t...

Страница 5: ...h the infrared sensor with a finger or breathe on it The probe is the most delicate part of the thermometer Use with care when cleaning the lens to avoid damage DO NOT attempt to open or disassemble the product DO NOT let children play with the product it is not a toy Contains small parts that may cause a choking hazard Keep children away from packaging product and its components In an emergency c...

Страница 6: ... 1 x Packet of Probe Covers 15pc 1 x User Manual Warranty Information 2 x AA Batteries Before Using the Ear Thermometer Carefully remove all packaging materials and keep the enclosed documents in a safe place Keep the packaging product and its components out of reach of children Check the product is intact DO NOT use if broken If any abnormalities are detected please contact our service centre In ...

Страница 7: ...ver F SET Button G Probe Eject Button LCD Display D F E G B A C Date Time Unit sign C or F Temperature indicator Temperature result Low voltage sign Memory mode Sound symbol Object Temperature mode Ear Temperature mode Probe Cover symbol Welcare_WET01_EarTherm_IM indd 7 Welcare_WET01_EarTherm_IM indd 7 3 12 21 2 21 3 12 21 2 21 ...

Страница 8: ... for 3 seconds with a happy face when a measurement is completed with a reading less than 37 8 C but higher than 37 3 C High Fever The display will light up RED for 3 seconds with a sad face when a measurement is completed with a reading equal to or higher than 37 8 C Object mode Note The display screen will light up GREEN for 3 seconds when the unit is ready for measurement Once the OBJECT measur...

Страница 9: ...ening of medical conditions Switching between C and F C will appear on the display Press Button to select and choose the desired mesurement unit either C or F and press SET button Setting the Time Date The 24H time format figure will appear on the display The device can display the time as either an AM PM 12 hour or a 24 00 24 hour format Set the Time Format Press and release Button to select the ...

Страница 10: ...il the correct year appears After the year is set press SET Button the Month figure will appear Set the month Press and release the Button to advance one month until the correct month appears After the month is set press SET Button the Date figure is flashing automatically Set the day Press and release the Button to advance one day until the correct month appears After the day is set press SET But...

Страница 11: ...itch between Ear Mode and Object Mode The chosen mode icon will appear on the screen Object Mode Ear Mode Note The unit will always default to Ear Mode when turned on Note The unit will only switch to Object mode if NO probe cover is present Reset Parameter Format Press the Button to turn unit OFF Press SET button for 5 seconds to turn the unit ON Follow the Using the Ear Thermometer instructions ...

Страница 12: ...asurement Place a new probe cover over the probe firmly making sure the probe and the inner portion of the probe cover fits securely and the probe is covering the connection ring and RED Micro Switch see below illustration When the probe cover is installed correctly the icon will appear on the LCD screen To release the probe cover gently press the Probe Eject Button RED Micro Switch Probe Eject Bu...

Страница 13: ...D screen The unit will always default to Ear Mode when turned ON Note If error message appears the thermometer is not ready for measurement Gently pull the ear back to straighten the ear canal and snugly position the probe into the ear canal aiming towards the membrane of the eardrum to obtain an accurate reading When using the thermometer on a person OVER the age of 1 year old pull the ear back m...

Страница 14: ...rm If the thermometer detects a body temperature 37 8 C or above three short beeps sound and will follow with one long beep to warn the user of potential fever Measuring Object Temperature Eject the probe cover Press the Button Press and hold the SET Button for approximately 3 seconds to switch to Object Mode Aim the thermometer at the centre of the object you want to measure with a distance of 1 ...

Страница 15: ...that correspond with a number 1 30 left The number 1 reflects the most recent reading while the number 30 reveals the oldest reading stored in memory The thermometer will automatically memorize the last 30 temperature readings Each memory also records the measurement date time and mode icon When in Memory Mode the user can press the TEST Button to take new measurements To turn the unit OFF when in...

Страница 16: ...eries as shown below figure E Clip the battery cover back on NOTE To protect the environment discard the used batteries in accordance with local regulations Replacement Accessories Replacement accessories can be purchased online or by contacting us directly Australia 1800 807 464 www welcare com au New Zealand 0800 523 583 www jadavey co nz Part Code Part Name WET01 60 Welcare Ear Thermometer Prob...

Страница 17: ...ime before taking temperatures Use a soft dry cloth to clean the thermometer display and exterior Do not submerge the thermometer in water Periodic cleaning and disinfection of the device following use to prevent patient cross infection Use a soft cloth slightly moistened with a 75 isopropyl alcohol solution to disinfect the thermometer and probe Do not use abrasive cleaners The probe is the most ...

Страница 18: ...is stored in a significantly different environment than testing location place it in the testing location for approximately 30 minutes prior to use Er 3 The thermometer is placed incorrectly in the ear Read Measuring Ear Temperature on page 13 Take a new temperature measurement Er 4 The thermometer showing a rapid ambient temperature change Allow the thermometer to rest in a room for at least 30 m...

Страница 19: ...y then check the integrity of the probe cover and take a new temperature measurement Lo In Ear mode Temperature taken is lower than 32 0 C 89 6 F In Object mode Temperature taken is lower than 0 C 32 F The thermometer s battery is low Replace with two new alkaline batteries size AA If the problem is not solved by the troubleshooting guide or you have any queries please contact our service centre I...

Страница 20: ... Laboratory Accuracy Ear mode 0 2 C 0 4 F during 35 5 C 42 0 C 95 9 F 107 6 F at 15 C 35 C 59 0 F 95 0 F operating temperature range 0 3 C 0 5 C for other measuring and operating temperature range Object mode 4 or 2 C 4 F whichever is greater Display Resolution 0 1 C or 0 1 F Measure Time Approximately one second Clinical Accuracy Clinical bias 0 08 C 0 14 F Clinical repeatability 0 13 C 0 23 F Li...

Страница 21: ... 6000 readings IP Rating IP22 Operating Conditions Temperature 10 C to 40 C Humidity 15 85 non condensing Atmospheric Pressure 700hPa 1060hPa Storage Transport Conditions Temperature 25 C to 55 C Humidity 15 95 non condensing Atmospheric Pressure 700hPa 1060hPa 0197 Welcare_WET01_EarTherm_IM indd 21 Welcare_WET01_EarTherm_IM indd 21 3 12 21 2 21 3 12 21 2 21 ...

Страница 22: ...dies in respect of the product which the consumer has under the trade practices act and other state or territory laws in Australia and New Zealand Our goods come with guarantees that cannot be excluded under Australian and New Zealand Consumer Law You are entitled to a replacement or refund for a major failure and for compensation for any other reasonably foreseeable loss or damage You are also en...

Страница 23: ...r contact details are still current and include a brief description of the problem you are experiencing together with your purchase receipt 2 Include the unit with all product components in your package This is extremely important so our repair technicians can make the correct diagnosis with any problems 3 Please return the product and warranty form at your cost to IN AUSTRALIA J A DAVEY PTY LTD W...

Страница 24: ...24 Welcare Ear Thermometer are_WET01_EarTherm_IM indd 24 are_WET01_EarTherm_IM indd 24 3 12 21 2 21 pm 3 12 21 2 21 pm ...

Страница 25: ...R WARRANTY NAME ADDRESS POSTCODE DAYTIME PHONE EMAIL MODEL DATE OF PURCHASE PLEASE ATTACH PROOF OF PURCHASE RETAILER S NAME RETAILER S ADDRESS POSTCODE BRIEF DESCRIPTION OF PROBLEM YOU ARE EXPERIENCING WARRANTY IS VOID UNLESS THE ABOVE INFORMATION IS COMPLETED AND CORRECT Welcare_WET01_EarTherm_IM indd 25 Welcare_WET01_EarTherm_IM indd 25 3 12 21 2 21 3 12 21 2 21 ...

Страница 26: ...26 Welcare Ear Thermometer are_WET01_EarTherm_IM indd 26 are_WET01_EarTherm_IM indd 26 3 12 21 2 21 pm 3 12 21 2 21 pm ...

Страница 27: ...Instruction Manual Warranty Information 27 Welcare_WET01_EarTherm_IM indd 27 Welcare_WET01_EarTherm_IM indd 27 3 12 21 2 21 3 12 21 2 21 ...

Страница 28: ...H HEALTHCARE CO LTD No 365 Wuzhou Road Yuhang Economic Development Zone 311100 Hangzhou China Manufactured for IN AUSTRALIA Australia J A DAVEY PTY LTD 626 Lorimer Street Port Melbourne VIC 3207 Telephone 1800 807 464 Website www jadavey com au IN NEW ZEALAND New Zealand J A DAVEY LTD C Healthcare Logistics 58 Richard Pearse Drive Mangere Auckland 2022 Telephone 0800 523 583 Website www jadavey co...

Отзывы: