26
English
In the case of relatively long, uninterrupted
use, the flow initially set gradually decreases
as cylinder pressure drops. To guarantee an
accurate dosage, we recommend looking at
the setting regularly and adjusting the flow if
necessary.
Multistep pressure reducer (Fine)
Set the multistep adjuster
value.
5.3 After operation
Important
: Never empty the oxygen cylinder
completely. Always ensure that there is a cer-
tain residual pressure in the cylinder when you
return it for filling, as this prevents moist ambi-
ent air from entering the cylinder and causing
corrosion.
1. Check the remaining oxygen content on
the cylinder in good time, e.g. when the
pressure drops below 50 bar, to ensure
a sufficiently long operating time.
Содержание OXYWAY Fast I
Страница 2: ...English 2 Français 3 Italiano 3 Nederlands 56 Español 57 Polski 57 ...
Страница 112: ......
Страница 113: ......