22
Operation
EN
2. Press the Softstart key (+) or the humidifier key (-) until the desired setting
appears in
the display:
3. To save the setting and exit the patient menu, press the on/off key/ESC or wait until the
menu exits automatically.
5.4 Softstart setting
To facilitate falling asleep, specialist staff can set a pressure which may deviate from the
optimum therapy pressure. This pressure is administered only for a certain time.
If the Softstart function is activated on your device, the time can be selected in 5-minute
increments up to a maximum of 30 minutes.
Set time
1. Call up the patient menu and scroll to the setting for time.
or:
with the device switched on: keep the Softstart key depressed until the current setting
appears.
2. To change the time, press the Softstart key (+) or the humidifier key (-) several times until
the desired time is shown.
3. To save the setting and exit the menu, press the on/off key/ESC or wait until the menu
exits automatically.
Softstart on/off
Status:
device switched on.
– To switch Softstart on or off, press the Softstart key.
When Softstart is switched on, the remaining Softstart time is shown in the display.
AVto
on
Automatic system activated
As soon as a patient breathes into the mask (pressure > 0.5 hPa), the
SOMNO
vent
CR switches on automatically.
When the mask is removed, the device switches off automatically
after 5 seconds.
AVto
0FF
Automatic system deactivated
The device can only be switched on by pressing the on/off key.
If the device is not used for 15 minutes, it switches off automatically.