Set up device
15
EN
4. Set up device
4.1 Set up and connect the SOMNO
vent
CR
Location
– Set up the device on a flat surface, e.g. on a bedside cabinet or on the floor next to the
bed.
– Maintain a distance of at least 5 cm between the wall and the rear of the device, as the
air inlet is at the rear of the device.
Caution!
Do not cover the device with blankets etc.. Do not operate it close to a radiator or in
direct sunlight. The device could overheat. This may lead to inadequate therapy and
damage to the device.
Power supply
1. Use the power cord to connect the power connector of
the SOMNO
vent
CR to a socket.
SOMNO
vent
CR automatically detects whether 115 V
or 230 V are connected and switches automatically.
Exception:
the SOMNO
vent
CR 230 V (WM 23470)
can only operate with a voltage of 230 V.
In the display, "
0
" appears for standby.
2. If the SOMNO
vent
CR is to be operated at a voltage of
12 or 24 V DC, connect the inverter to a cigarette
lighter socket.
The options for doing this are either:
• inverter 12 V WM 24131
• inverter 24 V WM 24132
Use the power cord to connect the power connector of
the SOMNO
vent
CR to the power supply socket of the
inverter.