2
Input Calibration points
Range
Values
Notes
Volts/Thermocouple
10 mV, 20 mV,
50 mV, 80 mV,
200 mV, 1 V
If the internal switches are set to 10 V, you must supply
a signal ten times the displayed value, e.g. 10 V for the
1V point.
mA
20 mA
RTD/Resistance
100 Ω, 200 Ω,
500 Ω, 800 Ω,
2 kΩ, 10 kΩ
Use four wire mode
Individual calibration
As selected
Allows you to recalibrate any of the points above
Fine adjust
Current input
Fine tune the adjustment for the range you have selected.
All
All ranges
Complete calibration of all points. Internal switches must
be in the 1V position.
Input calibration
1. Save the current calibration to disk (select ‘Save Calibration data’
from the ‘Device’ menu).
2. Select ‘Calibrate Input’ from the ‘Device’ menu.
. Enter the calibration password (Default is 101).
. Choose the calibration required (see Input calibration points table for
guidance).
5. Supply the signals requested by the software and follow the
instructions on screen. Press the ‘Done’ button when finished.
Содержание ITXPlus
Страница 2: ......
Страница 3: ...de ITXPlus Inhaltsverzeichnis Betrieb 4 Installation 6 Konfiguration 8 Kalibrierung 12...
Страница 15: ...15 en ITXPlus Contents Operation 16 Installation 18 Setup 20 Calibration 23...
Страница 26: ...26 fr ITXPlus Table des mati res Fonctionnement 27 Installation 29 Configuration 31 Etalonnage 35...
Страница 38: ...38 it ITXPlus Indice Funzionamento 39 Installazione 41 Setup 43 Calibrazione 46...
Страница 49: ...49 es ITXPlus ndice Funcionamiento 50 Montaje 52 Configuraci n 54 Calibraci n 58...
Страница 61: ...61...
Страница 62: ...62...
Страница 63: ...63...