21
Changing the instrument setup
1. Open the Setup/Calibration software.
2. Enter your initials using the ‘Enter Initials’ command from the
‘Device’ menu (shortcut is F7).
. Press the ‘Get from instrument’ button at the bottom of the screen
(shortcut is F9).
. The screen will now show the instruments current configuration.
5. Save the current configuration to disk using the ‘Save as’ command
from the ‘File’ menu.
6. Setup the instrument to suit your application via the
input screen (press F5) and output screen (press F6).
7. Return to the main screen, check the details are correct and press
the ‘Send to Instrument’ button (shortcut F8).
Note: You are required to enter a password to change the instruments
setup. The default password is 100. Make a note of the instrument
password if you change it - otherwise you will have to send the
instrument back to us.
8. Save the changes to disk as a record of changes you have made.
Main screen
Fields
Description
ID Tag
User defined field (e.g., PT1015-2)
User reference
User defined field (e.g., Pump 1015 bearing)
Содержание ITXPlus
Страница 2: ......
Страница 3: ...de ITXPlus Inhaltsverzeichnis Betrieb 4 Installation 6 Konfiguration 8 Kalibrierung 12...
Страница 15: ...15 en ITXPlus Contents Operation 16 Installation 18 Setup 20 Calibration 23...
Страница 26: ...26 fr ITXPlus Table des mati res Fonctionnement 27 Installation 29 Configuration 31 Etalonnage 35...
Страница 38: ...38 it ITXPlus Indice Funzionamento 39 Installazione 41 Setup 43 Calibrazione 46...
Страница 49: ...49 es ITXPlus ndice Funcionamiento 50 Montaje 52 Configuraci n 54 Calibraci n 58...
Страница 61: ...61...
Страница 62: ...62...
Страница 63: ...63...