Connection of a TW156V
Î
Apply pneumatic pilot pressure to the connector.
Î
Press the valve button (Pos. 4) (
jaws open.
Î
Be careful to place the connector with actuated valve button
(Pos. 4) straight and fully up to the stop onto the test piece.
Doing so will prevent any damage to the seal on the test
piece and damage to the connector and ensure a correct
connection.
Attention:
Insert the connector carefully and not with
force onto the test piece.
Î
Release the valve button (Pos. 4). The connector tightens
itself to the test piece and creates a pressure-tight connec-
tion (
).
Attention:
Make sure that the clamping jaws on the test
piece grip correctly, that the connector is seated properly
(it must not wobble) and that the pressure-tight connec-
tion has been made.
Î
In addition, screw the test piece into the connector hand-tight with a
1
/
4
turn
).
Attention:
The red ring must no longer be visible (
). Only then is the connec-
tor connected correctly. Otherwise the connector may spring out of place and
medium may escape. If the red ring is still visible, then the connector is not
). Repeat the connecting procedure.
Î
Pull lightly on the connector in the direction opposite that of the connection to
make sure that the connector is firmly in place.
- The connector is now connected pressure-tight to the test piece.
Î
To start the testing procedure, apply pressure to media inlet ‘B1’.
Figure 7
Figure 6
Figure 9
Figure 10
Red ring
Figure 8
MD-10021-L51-R2.1.0-03
Page 19
Operating instructions