HOT & COLD HAMMER
FAQ
Skin scrubber
Notes
1.
Press “STATE” to select between cold or hot
function.
2.
According to the needs, adjust the output
temperature by pressing “ENERGY” and desired time
of operation by pressing “TIMER”.
3.
Press “START/PAUSE” to start working.
•
For first time usage, function will start 5 minutes
after all set up.
•
After the first treatment, it needs 3 minutes before
the next Cold or Hot treatment.
•
Best working temperature condition is 25C.
•
Do not block the air tank while using this function, if
not steaming water will be blocked inside the Cold &
Hot hammer, and may induce short circuit.
•
Even after the “START/PAUSE” button has been
turned on to “START”, but need to check if the hammer
has been switched to “o“, if still in “-”, hammer would
not start working.
The probe suddenly stops emitting:
•
If the probe is separated from the skin, it will stop
emitting after 5 seconds. This is a safety feature and
should not be mistaken for a system fault.
•
Touch the skin with the probe and it will start
emitting again after 5 seconds.
Cautions
•
The 2 sides of Cold & Hot Hammer are made of
Aluminum Alloy which is not hard, avoid being knocked
on them.
•
If there is cracks on the 2 sides of hammer, replace
with new one
•
Be careful of the elastic wire working extension, do
not over pull it.
Ultrasound
Check the following if the work unit switches on but
the probes do not emit.
•
Put a drop of water on one probe.
•
Switch on the unit and slowly increase the output
energy.
•
If the water starts to move and evaporate, the unit
and the probes are working properly. If not, the probes
are not working properly.
The unit and the probes are working properly but the
customer doesn’t feel any effects.
•
Make sure enough conductive gel has been applied
to the treatment area.
Diamond dermabrasion
If the suction effect is reduced:
•
Replace the metal and foam filters.