Chapter 4 - Remote Interface Reference
Confi guring the Sensor
4
•
When writing the
[Sensor Bias Confi guration]
command, Character
7 (+ or -) confi gures the instrument to manual or automatic mode.
A ‘+’ puts the sensor confi guration under manual control. A ‘-’ lets
the instrument automatically confi gure the sensor. When a ‘-’ is in
Character 7, Characters 6 through 0 are ignored when writing.
Character 6 determines the feedback polarity. A thermistor is a Negative
Temperature Coeffi cient Sensor. All others are Positive Temperature
Coeffi cient Sensors.
You can combine settings to integrate a unique sensor. For example, if
you need 11 mA bias current, you can set Character 4 and Character 2
to ‘1’ to produce an 11 mA bias current.
•
When reading the Sensor Bias Configuration, the Command Type
changes from a WRITE (2) to a READ (1):
Read
!000.00020
(Reads
[Sensor Bias Confi guration]
)
This helps you determine what sensor confi guration the instrument
is using.
93
Содержание LFI-3751
Страница 10: ...10 This page intentionally left blank ...
Страница 13: ...1 Chapter 1 Quick Start 13 1 ...
Страница 20: ...20 This page intentionally left blank ...
Страница 21: ...2 Chapter 2 Front Panel Operation 21 2 ...
Страница 34: ...34 This page intentionally left blank ...
Страница 54: ...54 This page intentionally left blank ...
Страница 68: ...68 This page intentionally left blank ...
Страница 69: ...3 Chapter 3 Rear Panel Operation 69 3 ...
Страница 75: ...3 4 Chapter 4 Remote Interface Reference 75 4 ...
Страница 88: ...88 This page intentionally left blank ...
Страница 127: ...5 Chapter 5 Specifications 127 5 ...
Страница 131: ...Appendix CAT 220 Cable Accessory Diagram 131 ...