
73
3
Analog Input BNC
Input a 0 to
±
10 V analog signal to adjust the setpoint remotely. The sum
of the front panel
SET T
setting and analog input determine the control
temperature. If not using the analog input, install the 50
Ω
BNC terminator
provided with the unit to minimize electrical noise and protect against
electro-static discharge. The transfer functions change according to the
sensor type and bias current.
Thermistors & RTDs (10
µ
A or 100
µ
A)
Thermistors & RTDs (1 mA or 10 mA)
LM335/35, AD590, Thermocouples
10 k
Ω
/ V
100
Ω
/ V
10°C / V
Analog Input Transfer Functions
Analog Input BNC
BNC Connector [Used with Analog Version only]
Chapter 3 - Rear Panel Operation
BNC Connector
•
Example: With a thermistor biased with 100
µ
A, if
SET R
is 12 k
Ω
, if you
input +0.2 V at the
Analog Input BNC
, the setpoint becomes:
12 k
Ω
+ (0.2 V) (10 k
Ω
/ V
)
= 14 k
Ω
•
RS-232 Command:
[MOD Gain Confi guration]
Содержание LFI-3751
Страница 10: ...10 This page intentionally left blank ...
Страница 13: ...1 Chapter 1 Quick Start 13 1 ...
Страница 20: ...20 This page intentionally left blank ...
Страница 21: ...2 Chapter 2 Front Panel Operation 21 2 ...
Страница 34: ...34 This page intentionally left blank ...
Страница 54: ...54 This page intentionally left blank ...
Страница 68: ...68 This page intentionally left blank ...
Страница 69: ...3 Chapter 3 Rear Panel Operation 69 3 ...
Страница 75: ...3 4 Chapter 4 Remote Interface Reference 75 4 ...
Страница 88: ...88 This page intentionally left blank ...
Страница 127: ...5 Chapter 5 Specifications 127 5 ...
Страница 131: ...Appendix CAT 220 Cable Accessory Diagram 131 ...