• Average interval:
used to select the time interval for calculation of the
average. The available intervals are 10 s, 15 s, 30 s, 1 min, 2 min, 5 min, 6
min, 10 min, 15 min and 30 min.
• Spatial average:
choice between a single temporal log or several spatial
measurements that will be represented in a column graphical chart (see 6.1.4).
POST-PROCESSING
This menu is only available with broadband
probes that have correction factors in
memory. It has 3 parameters:
• Enable correction:
allows correction of
the value measured at a given frequency
(Working frequency defined below) using
the factors measured during the latest
laboratory calibration.
• Limit value for results in %:
allows direct
reading of the field as a % of a limit defined
on the main measurement screen. This is a
software calculation taking the limit value in
V/m and comparing it with the reading from
the probe. You can select the lowest limit
(minimum value) of the regulation (within
the band width of the probe) or a specified
frequency (Working frequency defined
below).
Note:
If the minimum value is selected, no working frequency can be defined and
therefore the correction factors cannot be activated.
User’s Manual
Electromagnetic Field Meter SMP2
P.
23
Содержание SMP2
Страница 2: ......
Страница 49: ...Figure 11 Installation of the program User s Manual Electromagnetic Field Meter SMP2 P 47...
Страница 79: ...Screenshot 4A Screenshot 4B User s Manual Electromagnetic Field Meter SMP2 P 77...
Страница 95: ......
Страница 96: ...www wavecontrol com...
Страница 97: ...Safety Quality Service info wavecontrol com www wavecontrol com 34 93 320 80 55...