Page 21
Chapter 2 -
TnT-el
Portable Appliance
Test Instructions
[continued]
LEAKAGE TEST
Menu 2
More Information – Leakage Test
The Leakage Test allows the user to operate the DUT in normal operation conditions and
measure its Operating Leakage current. The displayed parameter is mA:
A predefined value for individual class types is programmed into your TnT-el & TnT+
appliance testers. These limits are set according to the AS/NZ3760. Should these values
change in future it can be simply altered in firmware. A Pass / Fail will also be displayed
at the end of the test.
The run time period can be adjusted (by 5sec increments) with the use of WinPATS Lite
program V2.0 and above. The factory default setting is 20sec. Once reprogrammed using
WinPATS Lite the TnT-el appliance tester can run independently with its new set run time.
The mA Display Range 0.0 to 22.0 mA.
Before operating ensure the equipment is firmly secured to eliminate the possibility
of causing injury or damage. This function turns the DUT on; any rotating devices
have clearance from body or object. Appliances may jump off benches or cause
serious damage if not secured properly.
CAUTION: This test turns the Device Under Test (DUT) ON.
10 Amp (Resistive) Load MAX. The TnT-el appliance tester
monitors for potential loads over 10 Amps and displays
accordingly.
To Stop Operation Press the Return /Enter key
at any time.
The unit will power down after 20 seconds as factory default.
!