LEAKAGE TEST 

Menu 2 

 

 

 

 

 

This test determines errors of leakage not otherwise detected in a normal insulation test.  

If  there  are  any  doubts  with  insulation  testing  of  the  equipment,  the  Standard  (AS/NZS 

3760 since 2001) allows for a leakage test to be carried out 

instead.

 The TnT-el & TnT+ 

appliance testers have been designed to perform these tests. The Leakage Test applies 

power to the Device Under Test (DUT) and measures any imbalance or leakage current. 

The  leakage  is  tested  to  the  limits of  the  class  types  specified  in  the  Standard  AS/NZS 

3760 ie. Class 1 = > 5mA as Fail.   

 

The Limit of imbalance measured on the TnT-el appliance tester will read well in excess of 

the  limits  set  in  mA.  However,  should  the  supply  circuit  be  protected  by  an  RCD  this 

device will trip any where between 10 to 30mA and trip the mains supply switch OFF. 

 

 

PROCEDURE: 

 

1.  Complete a VISUAL INSPECTION on the DUT (Device Under Test)  

[refer to page 13 for details – ‘Visual Inspection’].

  

 

IF THE DUT PASSED THE ‘VISUAL INSPECTION’, CONTINUE WITH THE FOLLOWING 

INSTRUCTIONS 

[if not, refer to page 13 ’IF RESULT IS A FAIL’]: 

 

2.  Press Return Key to view 

Menu 2.

  

3.  Press a Function Key (as follows) to start the test/s. Press F1, F2 or F3 (depending 

on the type of DUT you are testing). 

1.  F1 Earth  Leakage   

 

 

 Limit set to MAXIMUM then 5mA Fail

 

2.  F2 D. Insulated Leakage & Ext leads

  

Limit set to MAXIMUM than 2.5mA Fail

 

3.  F3 RCD Leakage 

CONFIRM SAFE TO CONTINUE? OK/QUIT. IF OK, DEVICE WILL TURN ON. 

ENSURE SAFE OPERATION! 

 

RESULTS ARE DISPLAYED 

4.  Read the result and record appropriately.  

5.  Unplug the DUT  

6.  Press the Enter key until you return to the Main Menu. 

 

IF THE RESULT OF THE TEST WAS A ‘PASS’  - Tag the DUT with the appropriate tag, 

including ‘Next Test Due’ date and return to service.    

IF THE RESULT OF THE TEST WAS A ‘FAIL’ - Tag the DUT with a DANGER TAG AND 

REMOVE THE DUT FROM SERVICE.  

Note: Test will PASS even if DUT is not plugged in. 

The Leakage Test is an alternate method to performing insulation resistance 

tests. Leakage testing is a major function of the TnT-el & Tnt+ models  

 

 

 

 

 

PLEASE NOTE: A VISUAL INSPECTION MUST BE CARRIED OUT 

BEFORE ANY OTHER TESTS.  

[refer to page 13 for more details] 

Содержание TNT+

Страница 1: ...08 8331 8892 Ph 07 3274 2466 Int 61 8 8331 8892 Int 61 7 3274 2466 Email admin wavecom com au Email sales nesco com au www wavecom com au www nesco com au Australian made to comply with the requiremen...

Страница 2: ...Manufacturers Calibration Certificate IEC Lead Earth Lead with Alligator Clip Soft Carry Case Manufacturers Recommendations For the Range of TnT Portable Appliance Testers Calibration The manufacturer...

Страница 3: ...ST 12 FIRST TEST VISUAL INSPECTION TEST 13 CHAPTER 1 TnT Portable Appliance Tester Instructions TnT Test Function Flow Chart 14 EARTHED APPLIANCE TEST 15 DOUBLE INSULATION TEST 16 EXT LEAD TEST 17 CHA...

Страница 4: ...gainst the consequences of the incorrect use of an electrical device We advise that appliance testing should be conducted by a competent and suitably trained person For maximum safety always ensure th...

Страница 5: ...ay not be exact to each specific model of TnT Appliance Tester Display Test Function F1 Test Function F2 Test Function F3 These buttons are NOT IN USE are reserved for future functions Supply Indicato...

Страница 6: ...e results for compliance with the Standard Workplace requirements 8 The knowledge to be able to correctly recommend the frequency of testing required Due to the potential hazards of electrical testing...

Страница 7: ...protective earthing conductor earth wire in the fixed wiring of the device Installation The protective earthing terminal of the equipment must be marked with the word earth or the symbol E or the symb...

Страница 8: ...In Class II equipment protection against electric shock does not rely on basic insulation only but has additional supplementary insulation such as double insulation or reinforced insulation provided t...

Страница 9: ...these instances the test voltage should be changed to 250V then retest If DUT still fails check with the DUT Operators Manual or an electrician Refer to page 13 for details Double Insulation Test 250...

Страница 10: ...se plug E G 5 pin Clipsal 56 series 20A CAUTION Ensure that these adaptors are used only for appliance testing To Change Test Voltage 250 500 VDC Select To Change the Insulation Test Voltage on all mo...

Страница 11: ...ected to the TnT el supply If working with a generator or inverter this is most likely to occur and you may need to consult an electrician before proceeding If both the A E A N green lights are on but...

Страница 12: ...you to check if the DUT is plugged in and switched on If the DUT is not plugged in and or recognised it may require a live test therefore making it necessary for the operator to carry out a full funct...

Страница 13: ...ds and or leads are effectively anchored glands and grommets intact 8 All covers guard are in place and secure as intended by the supplier manufacturer 9 All safety devices and systems are in good wor...

Страница 14: ...3Ext Lead Test F1 Earth Test Supply Mains Test NCNT Test Earth Bond Test Insulation Test Results Summary F2 D Ins Test Supply Mains Test NCNT Test Double Ins Test Results Summary F3 Ext Lead Test Sup...

Страница 15: ...ASS Tag the DUT with the appropriate tag including Next test Due date and return to service IF THE RESULT OF THE TEST WAS A FAIL Tag the DUT with a DANGER TAG AND REMOVE THE DUT FROM SERVICE In some s...

Страница 16: ...Main Menu IF THE RESULT OF THE TEST WAS A PASS Tag the DUT with the appropriate tag including Next test Due date and return to service IF THE RESULT OF THE TEST WAS A FAIL Tag the DUT with a DANGER TA...

Страница 17: ...RESULT OF THE TEST WAS A PASS Tag the DUT with the appropriate tag including Next test Due date and return to service IF THE RESULT OF THE TEST WAS A FAIL Tag the DUT with a DANGER TAG AND REMOVE THE...

Страница 18: ...A Results Summary F1 Earth Test Mains Supply Test NCNT Test Earth Bond Test Insulation Test Results Summary F2 D Ins Test Mains Supply Test NCNT Test Double Ins Test Results Summary F3 Ext Lead Test M...

Страница 19: ...Portable Appliance Test Instructions TESTS COMMON TO ALL MODELS Menu 1 EARTHED APPLIANCE TEST refer pg 15 DOUBLE INSULATION TEST refer pg 16 EXT LEAD TEST refer pg 17 Please note Menu 1 is common to a...

Страница 20: ...sual Inspection IF THE DUT PASSED THE VISUAL INSPECTION CONTINUE WITH THE FOLLOWING INSTRUCTIONS if not refer to page 13 IF RESULT IS A FAIL 2 Press Return Key to view Menu 2 3 Press a Function Key as...

Страница 21: ...with the use of WinPATS Lite program V2 0 and above The factory default setting is 20sec Once reprogrammed using WinPATS Lite the TnT el appliance tester can run independently with its new set run ti...

Страница 22: ...PF LEAKAGE TEST F1 Earthed Devices F2 D Ins Ex Lead To return to MENU 2 Press RETURN EARTH LEAKAGE TEST Continue if Safe OK QUIT LEAKAGE RESULTS PASS 5 0 mA 0 0 mA Pass ESC EMERGENCY STOP ABORT To ret...

Страница 23: ...ortable Appliance Test Instructions TESTS COMMON TO ALL MODELS Menu 1 EARTHED APPLIANCE TEST refer pg 15 DOUBLE INSULATION TEST refer pg 16 EXT LEAD TEST refer pg 17 Please note Menu 1 is common to al...

Страница 24: ...the Extension lead Power board 7 Press the Enter key to return to the Main Menu IF THE RESULT OF THE TEST WAS A PASS Tag the DUT with the appropriate tag including Next test Due date and return to ser...

Страница 25: ...n the appliance on and measure its performance as a digital wattmeter Displayed parameters are Volts AC Current Volt Amps Power factor Watts Before operating Ensure the equipment firmly secured and is...

Страница 26: ...ng large areas in the work place monitored by the switchboard RCD it is suggested that an RCD isolation transformer be used TnT ISOT These are designed specifically for the purposes of field RCD tripp...

Страница 27: ...isplay 3 functions Power Test Leakage Test RCD Test To enter RCD menu press and release F3 button RCD sub menu This will display a further 2 options F1 RCD Trip Time F2 RCD Trip Current CAUTION Pressi...

Страница 28: ...0mA RCD tester from 5mA to max 500mA output 30mA X 5 0 150mA These tests should result in no trip trip fast trip times respectively F1 0 deg F2 180 deg Change Test mA F3 Change test mA This allows the...

Страница 29: ...s button sets the selected current for the next trip time test The TnT will then return to the current trip time test screen The displayed trip time is in milliseconds This is the time taken for the R...

Страница 30: ...inconsistent in performance The TnT has a nominal leakage current of 2mA which should be added to the result of test E g if RCD tripped at 22mA 2mA TnT 24mA trip current RCD Trip Current Press F2 for...

Страница 31: ...se protected 500Vac 10A M205 X Models Only 20A Thermal overload protected manual reset Power Test for TnT model only Voltage 220 270 Vac 5 of reading Current 0 10 A 5 of reading X Models Only 0 20A Lo...

Страница 32: ...TS Software Upgrade Part No WCM WinPATUP dependant on specific upgrade Blue Tooth Interface Part No WCM BTHT WinPATS Scanner 232 Part No WCM 1200 Scanner RS232 WinPATS 1200 Scanner USB Part No WCM 120...

Страница 33: ...lia Ph 03 6233 7831 Office of the Chief Electrical Inspector Ministry of Commerce New Zealand Energy and Resources Division P O Box 1473 New Zealand 6015 Wellington Office of Energy WA Technical and S...

Страница 34: ...Page 34...

Страница 35: ...Page 35...

Страница 36: ...Page 36 Ph 08 8331 8892 Ph 07 3274 2466 Int 61 8 8331 8892 Int 61 7 3274 2466 Email admin wavecom com au Email sales nesco com au www wavecom com au www nesco com au...

Отзывы: