Page 14 of 18
Annual Preventive Maintenance
Note: If using pliers to remove
NEW
membrane for any reason, protect end of membrane with
cloth, to prevent teeth marks which could result in leaks.
Perform quarterly preventive maintenance.
Include replacement of O-Rings.
Replace RO membrane by
loosen retaining clamp at top
of stainless steel membrane
vessel
.
Carefully pry loose PVC end
cap from stainless steel
vessel.
Pull membrane from vessel using pliers i
necessary and discard. Take note of
location of black BRINE SEAL located
approx. 1/2” from one end of membrane,
so the new membrane will be installed in
the proper direction.