AccUView LED (1/2020)
Rev 2
21
7.8 Speed of Response
The speed of response for both displayed and output values can be adjusted in this menu.
The factory setting is 3 however, 10 response speeds are available. Select the desired
speed of response using the
and
buttons. Press the
↵
button to accept it.
Please note that the response is simply an average of the previous readings combined with
the current reading. If the setting is 1, only current readings are displayed. If the setting is
10, the last 9 readings are averaged with the current reading.
7.9 LCD Backlight Brightness
The LCD backlight brightness may need to be adjusted. This is of particular interest if
multiple instruments are located in the same area and it is desired for the entire group to
have the same appearance. Ten levels are available. The factory brightness is 8.
Change the brightness by pressing the
or
button. When the desired brightness has
been selected, press the
↵
button.
7.10 Ultrasonic Cleaning
This allows for a selection menu to turn
OFF
the ultrasonic cleaning function if desired.
The default mode is
On
. Make a selection using the
and
buttons then press the
↵
button.