AccUView LED (1/2020)
Rev 2
16
7.2.1 Setting the 4 mA Level
Select the %T or ABS level to assign to the
4MA
using the
and
buttons.
The factory setting is 90%T or.
Once the desired level has been set, press the
↵
button to accept it.
The factory default setting for %T is 90%T and the for ABS is 0
The 4MA cannot be set higher than the 20 MA level.
7.2.2 Setting the 20 mA Level
The next, prompt will be the %T or ABS level assigned to the 20 mA output level (
20MA)
on the lower row of the LCD display). Select the %T level to assign to the
20MA
using
the
and
buttons. Once the desired level has been set, press the
↵
button to accept it.
The factory default setting for %T is 100%T and for ABS is 2.
The 20 MA cannot be set lower than the 4 MA level.
7.2.3 Setting the Error Level
In case of an error in the AccUView LED, the 4-20 mA reading can be used to indicate a
problem by sending the output to either:
1.
4.00 mA,
2.
2.00 mA
3.
0 mA
4.
OFF. In the case of OFF, the 4-20mA is unaffected by any error condition.
The factory default setting is OFF. Select the desired ERLV by using the
and
buttons then press the
button to accept the desired error response.