AccUView LED (1/2020)
Rev 2
15
7.0 Instrument Configuration (CONFIG mode)
The instrument has been designed to provide the ability to customize the instrument
according to needs at any time during normal operation. This mode has been split into
sub-menus to facilitate instrument configuration. This section describes how to use each
of the sub-menus to configure the instrument. While in the configuration mode, the
instrument has a time-out feature that automatically returns the system operation to the
AUTO
mode after a fifteen (15) minute period.
Enter the
CONFIG
mode of the instrument by pressing the
MODE/EXIT
button until the
arrow beside
CONFIG
is illuminated, then press the
↵
button.
To exit the CONFIG mode, press the
MODE/EXIT
button.
7.1 Selecting the Output (O/P)
The first configuration selection is the
O/P
. The selections are:
1.
4-20
for the 4-20 mA output
2.
485
for the RS-485 and modbus
3.
OFF
if no outputs are required.
Select the desired output by using the
and
buttons. Once the desired output has
been set, press the
↵
button to accept it. The next prompts will depend on the output
selected.
7.2 Setting the 4-20 mA
If the 4-20 mA output was turned on, prompts to set the 4mA
(4MA)
and 20mA
(20MA)
%T or ABS limits levels will be displayed. There is also a menu to adjust the error level
(ERLV
). The first prompt will be the %T or ABS limit assigned to the 4 mA output level:
.