If the message constantly repeats after several pumphead installations, then there may be
a leak detection sensor failure. Please contact Watson-Marlow for repair.
If you need to bypass a faulty leak detection sensor then you can do so by pressing
IGNORE
. If the power is cycled, the leak detection sensor will be switched back on.
If the leak detector is
IGNORED
then the leak detection off icon will be displayed on the
HOME
screens.
23.2
Error codes
Note:
For qdos remote refer to "Error Indication (Remote only)" on the facing page
If an internal error occurs, an error screen with a red background is displayed.
Note:
Signal out of range
and
Leak detected
error screens report the nature of an external
condition. They do not flash.
Error
code
Error condition
Suggested action
Er 0
FRAM write error
Attempt to reset by switching power OFF / ON. Or seek
support
Er 1
FRAM corruption
Attempt to reset by switching power OFF / ON. Or seek
support
Er 2
FLASH write error during drive
update
Attempt to reset by switching power OFF / ON. Or seek
support
Er 3
FLASH corruption
Attempt to reset by switching power OFF / ON. Or seek
support
Er 4
FRAM shadow error
Attempt to reset by switching power OFF / ON. Or seek
support
Er 9
Motor stalled
Stop pump immediately. Check pumphead and tube.
Power OFF / ON may reset. Or seek support
Er10
Tacho fault
Stop pump immediately. Power OFF / ON may reset. Or
seek support