6.
Prepare the cable screen(s) by twisting a
suitable length. The twisted length(s) shall
ideally be sleeved to prevent shorting.
7.
Secure the end of the cable screen to the
Faston receptacles on the spade connectors
provided.
8.
Push the bared conductor into the square hole
in the connector. When the conductor is fully
inserted, tighten the retaining screw to secure
it in place.
Ensure that multi strand
wires are terminated with a
crimp, suitable for the wire
diameter. Failure to do so
may result in electric shock.
9.
When all conductors are in position replace
the module cover.
10.
Using a 21mm spanner tighten the gland cap
to 2.5Nm to ensure a watertight seal.
Alternatively, tighten the gland by hand until
it is finger-tight and use a spanner to tighten
it further by one-half a turn.
Ensure that unused
openings on the module are
sealed using the blanking
plugs provided. Failure to do
so will result in loss of
ingress protection.
m-qdos-en-02
33